Induztr
Engineering Digest Document Viewer Related Models

Rugged. Reliable.

Contents

  1. Overview pp. 1-2
    Introduces rugged, reliable instrument performance and service value
  2. Overview p. 3
    Explains tuning performance and service options overview
  3. Services p. 4
    Describes Agilent CrossLab real insight and outcomes
  4. p. 3 table, 26 rows

Determination of low As levels in high nickel ores

Sample IDConcentration mg/LMean AbsRecovery
CAL ZERO0.0–0.0033
STANDARD 125.00.1717
STANDARD 250.00.3342
STANDARD 375.00.4805
STANDARD 4100.00.6186
50ppm Standard 249.10.3281
Blank0.20.0017
Sample 00113.30.0916
Sample 001Dup13.80.0950104%
50 ppm Standard 245.60.306491%
Sample 01311.90.0816
Sample 0145.70.0395
Sample 013 Dup11.50.079197%
Sample 014 Dup5.40.037295%
Sample 0170.10.0007
100 ppm Standard 492.60.579093%
Sample 02323.40.1604
Sample 02412.20.0841
Sample 0250.60.0043
Sample 0260.90.0065
100 ppm Standard 497.30.602997%
Sample 023 Dup23.60.1613101%
Sample 024 Dup12.00.082898%
Sample 024 Dup11.60.080195%
Sample 023 Dup23.90.1634102%
Blank–0.3–0.0019

Excellent reproducibility and good read back on the QC standards is achieved, even with < 1 ppm As in solution, eliminating the need for extraction.

Engineering Facts

14 facts · every claim linked to a page
Selection

The instrument is a standalone double beam AA suitable for remote sites where a simple, self-contained analyzer is required.

Confirms the 55B is a standalone double-beam unit, which matters when selecting an AA for remote or harsh locations without a PC.

Source
“ideal for remote sites requiring a simple, standalone, double beam AA that doesn’t compromise on performance”
Installation

The instrument can be connected directly to a LIMS via RS-232, so a serial link must be planned during installation.

RS-232 LIMS connectivity affects cabling and integration planning at the installation site.

Source
“Configured for direct LIMS connection via RS-232”
Operation

Wavelength and slit selection are fully automated, so no manual optical alignment is needed by the operator.

Reduces operator skill requirements and setup time, which is relevant when choosing an instrument for routine or remote operation.

Source
“No complex alignment. Wavelength and slit selection is fully automated”
Operation

Two lamp positions allow one lamp to be prewarmed while the other is in use, improving accuracy and throughput for sequential elements.

Affects multi-element workflow planning and lamp configuration at commissioning.

Source
“Two lamp positions enable operation while prewarming the lamp for the next element”
Maintenance

Sealed optics with quartz overcoated mirrors protect against dust and vapors, supporting long-term stable performance in harsh environments.

Sealed optics reduce maintenance frequency and are a key selection criterion for dusty or corrosive lab environments.

Source
“Sealed optics with quartz overcoated mirrors provide protection from dust or vapors and ensure long term, stable performance”
Installation

An internal air purge maintains positive clean-air flow through the instrument, which requires a clean air supply to be provided at installation.

The air purge requires a clean air source and plumbing, which must be planned during site installation.

Source
“Enhance protection using the internal air purge to maintain a positive flow of clean air through the instrument”
Operation

The industrial LCD screen and dedicated keypad are chosen for rugged lab conditions instead of touch screens.

Affects operator interface reliability and suitability for harsh environments when selecting the instrument.

Source
“The industrial LCD screen and dedicated keypad are suitable for rugged laboratory conditions, unlike failure-prone touch screens”
Performance

The Mark 7 atomization system uses externally adjustable impact beads to tune performance for complex matrices.

External adjustment allows optimization for difficult samples without disassembly, which matters for method development and operation.

Source
“Tune performance using the externally adjustable impact bead”
Performance

Typical sensitivity is greater than 0.9 Abs from 5 mg/L Cu, giving a benchmark for expected signal response.

Provides a performance benchmark to verify instrument sensitivity during acceptance testing.

Source
“Achieve high sensitivity—typically > 0.9 Abs from 5 mg/L Cu”
Performance

Typical precision is better than 0.5% RSD from 10 five-second readings, setting an expectation for repeatability.

Gives a repeatability specification to check during installation and performance verification.

Source
“Optimize precision—typically < 0.5% RSD from 10 five second readings”
Operation

Removable twin-headed mixing paddles ensure thorough mixing and a superfine aerosol to reduce interferences in complex samples.

Affects atomization quality and interference handling for complex matrices, relevant to method setup.

Source
“The removable twin headed mixing paddles ensure thorough mixing and a superfine aerosol”
Maintenance

The contoured burner design resists blockage even with difficult samples, reducing burner maintenance.

Lower blockage risk reduces downtime and maintenance effort with difficult sample types.

Source
“Minimize burner blockage—the contoured design provides outstanding resistance to blockage, even with difficult samples”
Selection

Inert components make the system compatible with high acid and organic matrices, which affects material selection for the sample path.

Compatibility with aggressive matrices is a key selection criterion for labs analyzing acids or organics.

Source
“Inert components provide compatibility with high acid and organic matrices”
Performance

Deuterium background correction provides accurate correction with a fast 2 ms response time.

Fast background correction is important for accurate results with complex matrices and affects method performance.

Source
“Provides accurate correction with a fast 2 ms response time”