Agilent Flame Atomic Absorption Spectrometer User's Guide

Contents

  1. pp. 3-10
    Table of contents and page listings
  2. Warning pp. 11-32
    Safety warnings for gases, flames, radiation and perchloric acid
  3. Overview pp. 33-36
    Instrument introduction and environmental specifications
  4. Overview pp. 37-50
    Overview of instrument components, accessories and connections
  5. Overview pp. 51-68
    Display screen, keypad and interface navigation
  6. Procedure pp. 69-92
    Method development, flame lighting, calibration and measurement procedures
  7. Procedure pp. 93-130
    Maintenance schedule, cleaning, nebulizer and burner servicing
  8. Ordering pp. 131-136
    Spare parts and miscellaneous component listings
  9. Procedure pp. 137-154
    Troubleshooting tables and error message descriptions
  10. p. 131 table, 12 rows
  11. p. 35 table, 4 rows
  12. p. 140 table, 3 rows

Figure 94. Nebulizer components

№Component
1Thimble, capillary adjuster
2Locking ring
3Clamp bush guide
4Ertalyte bush bearing
5Capillary assembly
6O-ring 1/32 in. ID x 3/32 in. OD x 1/32 in. nitrile
7Spring
8Capillary guide
9Venturi
10O-ring 3/16 in. ID x 5/16 in. OD x 1/16 in. nitrile
O-ring 3/16 in. ID x 5/16 in. OD x 1/16 in. organics
11Impact bead adjuster screw

Пункт 10 включает две позиции: nitrile и organics

Table 2. Gas supply requirements

C2H2 (Acetylene)AirN2O (Nitrous Oxide)Air purge
Instrument grade >99.0% pureClean, dry, free of oilInstrument grade >99.5% pureMust be clean & dry. Air filter to be used.
Allowed range65-100 kPa (9.5-14.5 psi)245-455 kPa (35-65 psi)245-455 kPa (35-65 psi)245-455 kPa (35-65 psi)
Recommended75 kPa (11 psi)350 kPa (50 psi)350 kPa (50 psi)350 kPa (50 psi)
Normal flow rate (L/min)0-1013.5-2011-1610

Other gas connections — Sample compartment: Push-on Air/N2O connector for burner; Push-on C2H2 connector for burner

Table 11. Severity Codes

CodeMeaningExplanation
IInformationThis type of message provides information only, such as prompting the user to perform some action.
WWarning.Provides information about a system difficulty, or potential difficulty, which the user should be aware of.
EErrorA message relating to a condition that will terminate the current operation. The system will continue to run, but the operation will have failed.

Engineering Facts

14 facts · every claim linked to a page
Selection

Only air, nitrous oxide, and acetylene may be used for flame operation; oxygen or oxygen-enriched air must never be used as oxidant.

Determines which gas supply and regulators must be specified and installed for the site.

Source
“This spectrometer is to be used only with air, nitrous oxide, and acetylene for flame operation.”
Selection

Only Mark VI A or Mark 7 burners are compatible with this system; older burner versions must not be installed.

Prevents ordering or fitting an incompatible burner during system build.

Source
“Only Mark VI A or Mark 7 burners should be used with this system. Previous versions should not be installed.”
Selection

Burner slot width limits define when a burner must be replaced rather than refurbished.

Sets the wear criterion for burner replacement decisions and spare-part planning.

Source
“If the width of the burner slot exceeds 0.47 mm (0.0185 in) for a nitrous-oxide/acetylene burner or 0.54 mm (0.021 in) for an air/acetylene burner, the burner cannot be refurbished and must be replaced with a new one.”
Installation

The instrument must not be connected to mains until the operating voltage is confirmed correct for the actual outlet.

Prevents fire and shock hazard from wrong voltage setting or miswired outlet at commissioning.

Source
“Do not connect the instrument to the mains power supply until you have made sure that the operating voltage is correctly set for the mains power supply in the specific outlet in your laboratory to which the equipment will be connected.”
Installation

A grounded 3-wire outlet adequately rated for the load is mandatory.

Defines the electrical supply requirement the lab must provide before installation.

Source
“Always use a 3-wire outlet with ground connection which is adequately rated for the load.”
Installation

The chimney must sit about 100 mm from the extraction hood, with more distance for nitrous-oxide flames.

Governs exhaust hood positioning and clearance during site setup.

Source
“The chimney should be approximately 100mm (4in) from the extraction hood of your exhaust system. This distance should be increased when operating a nitrous-oxide flame due to the higher flame height.”
Installation

The sample compartment front panel must be securely fitted during operation.

Affects assembly checks and safe operation after installation or maintenance.

Source
“Ensure the sample compartment front panel is securely fitted on the spectrometer during operation.”
Operation

The liquid trap must be filled with the same solution as the sample/standard matrix.

Required setup step that affects both safety (gas seal) and analytical accuracy.

Source
“The liquid trap must be filled with the same solution as the matrix used for the standards and samples.”
Operation

The drain tube must slope downward continuously to the drain vessel and its end must stay above the liquid level.

Prevents pressure surges in the spray chamber that can cause flashback.

Source
“The tubing must slope downwards all the way to the drain vessel so that the waste liquid drains smoothly.”
Operation

Acetylene cylinder pressure must not drop below 700 kPa to avoid acetone carry-over.

Sets the cylinder replacement threshold and operating limit for the fuel gas supply.

Source
“If the gas is withdrawn too quickly, or the cylinder pressure is allowed to drop below 700 kPa, acetone may be drawn off in sufficient quantities to affect analytical performance, damage seals, O-rings and hoses, or even cause a flashback.”
Operation

A heater is needed on the N2O regulator to prevent icing during expansion.

Determines regulator accessory selection for nitrous-oxide operation.

Source
“Prevent freezing by using a heater on the N2O regulator on the supply cylinder.”
Maintenance

The D2 lamp needs replacement roughly every 1000 operating hours.

Sets the preventive maintenance interval and spare-lamp stocking.

Source
“You will need to replace the D2 lamp approximately every 1000 hours of operation.”
Maintenance

Gas hoses showing damage or deterioration must be replaced immediately by a trained Agilent representative.

Defines the hose inspection/replacement rule and who may perform it.

Source
“Hoses showing any signs of damage or deterioration must be replaced immediately. This should be done by a trained Agilent representative.”
Safety

UV exposure with the flame door open must be limited to 13 minutes per 8-hour period.

Sets an exposure limit that constrains maintenance and troubleshooting work with the door open.

Source
“Avoid exposure with the flame door open of more than 13 minutes over an 8 hour period.”