Covered Models & Document Scope

Quick reference for operating a 5973N MSD (Mass Selective Detector) with a 6890 GC using MSD Productivity ChemStation Software Rev. D.00.00; covers EI and CI MSD variants, autosampler (ALS) and manual injection, and Enhanced Mode software menus.

Section-by-Section Technical Breakdown

pp. 6-7
Instrument Control View

Describes the main ChemStation view for setting and monitoring instrument parameters. It explains the acquisition status, start/stop buttons, run time clock, and buttons for injector, inlets, columns, oven, auxiliary (GC/MSD interface), and MS parameters.

  • Start button is green when no run is in progress; click it, fill sample information, and click Start Run.
  • Stop button is red during Pre-Run, Run, or Post-Run mode.
  • Oven ready indicator: green square when GC is ready, red when not ready.
  • Aux button sets the GC/MSD interface temperature (usually Thermal Aux 2).
  • MS button edits MS acquisition parameters, selects tune file, or edits MS monitors.
pp. 8-15
Software Menus (Enhanced Mode)

Lists the menu commands available in the Top, Instrument Control, Diagnostics/Vacuum Control, Manual Tune, Data Analysis, EasyID, QEdit, Parametric Retrieval, Align GC, Review Peak Purity, Quick Screener, and RTLock views. These menus provide access to instrument control, tuning, data analysis, and vacuum diagnostics.

  • Diagnostics/Vacuum Control menu includes Pump Down, Vent, MS ON, MS OFF, Set RFPA, Air and Water Check, and MS Error Codes.
  • Manual Tune menu includes Autotune, Low Mass Autotune, Standard Spectra Tune, DFTPP Tune, BFB Tune, Target Tune, QuickTune, and calibration/ramp functions.
  • Data Analysis menu includes quantitation, library search, integration, and reporting tools.
  • RTLock menu provides retention time locking setup and calibration.
pp. 16
To vent (shut down) the MSD

Step-by-step procedure for safely venting the MSD, including switching off the gauge controller, turning off reagent gas for CI MSD, setting heaters to ambient, turning off power, and opening the vent valve 3/4 turn. Includes safety warnings for hydrogen carrier gas and hot surfaces.

  • Set GC/MSD interface heater and GC oven to ambient (25°C) before venting.
  • Turn vent valve knob counterclockwise only 3/4 turns or until hissing is heard.
  • On a CI MSD, the Gas Off light must be on when venting.
  • If using hydrogen carrier gas, carrier gas flow must be off before turning off MSD power.
  • Do not put ChemStation into Top view while MSD is vented; this turns on the interface heater.
pp. 17
To pump down (start up) the MSD

Preconditions and procedure for pumping down the MSD. Requires closed vent valve, secure vacuum seals, grounded power, conditioned column, carrier gas of at least 99.999% purity, and proper foreline exhaust. After pump down, waits for turbo pump to reach 95% and thermal equilibrium.

  • Carrier gas purity must be at least 99.999% with recommended traps.
  • If hydrogen is used, carrier gas flow must be off and front sideplate thumbscrew loosely fastened.
  • Rough pump gurgling noise should stop within a minute; continued noise indicates a large air leak.
  • Turbo pump should reach 80% within 10–15 minutes and eventually 95%.
  • After 'Ok to run' appears, wait two hours for thermal equilibrium before tuning or acquiring data.
pp. 18
To tune your MSD

Explains autotune and manual tune procedures. Autotune options include Autotune, Low Mass Autotune, Standard Spectra Tune, and QuickTune. Manual tune allows interactive adjustment of lens voltages and tuning masses, with profile and spectrum scans.

  • Always tune with the same GC oven temperature, column flow, and analyzer temperature used for data acquisition.
  • Autotune gives maximum sensitivity over the full scan range.
  • Standard Spectra Tune gives standard response over full scan range and may reduce sensitivity.
  • QuickTune adjusts peak width, mass assignment, and abundance without changing ion ratios.
  • Manual tune can provide greater sensitivity than autotune and allows ramping of individual parameters.
pp. 19
To acquire data

Describes GC setup for MSD use, autosampler injection, and manual injection. Includes verifying inlet/injection types, auxiliary channel 2 heater, and column configuration with MSD detector and vacuum outlet.

  • Select Instrument / Inlet/Injection Types, choose injection source, and select Use MS checkbox.
  • Verify Aux channel 2 heater is on and set to desired temperature with MSD selected as Type.
  • Verify column detector is MSD and Vacuum is selected for Outlet psi.
  • For autosampler, specify unique data file name and vial position (1–100).
  • For manual injection, press Prep Run key on GC keypad before starting run.
pp. 20
To analyze MS data

Covers loading data files, integrating chromatograms, selecting spectra, zooming, averaging spectra, adding/subtracting spectra, and subtracting background. These operations are performed in Data Analysis view.

  • Load data file via File / Load; chromatogram appears in window [2].
  • Integrate via Chromatogram / Integrate; select integrator first if needed.
  • Double-click right mouse button on chromatogram to select a spectrum.
  • Subtract Background (BSB) subtracts selected spectrum from every scan and stores result in BSB subdirectory.
  • Zoom out by double-clicking left mouse button in zoomed window.
pp. 21
To use spectral libraries

Procedures for integrating and searching peaks against spectral libraries, and for searching a single spectrum. Library search reports can be sent to screen, printer, or file.

  • Select Spectrum / Library Search Report to integrate TIC and search each peak.
  • Choose Summary or Detailed report format and destinations (Screen, Printer, File).
  • Select spectrum from each peak: Apex, Apex - Start of Peak, Apex - Background at time, or Peak Average.
  • For single spectrum search, double-click right mouse button in spectrum window.
  • Statistics button shows quality of each library hit.
pp. 22
To use retention time locking

Describes retention time locking (RTL) procedure to negate retention time changes from column changes. Involves collecting data at five inlet pressures (-20%, -10%, nominal, +10%, +20%) and generating a pressure/retention time curve.

  • RTL calibration runs are stored as RTLOCK1.D through RTLOCK5.D in method directory under RTLOCK folder.
  • Five runs are made automatically at -20%, -10%, nominal, +10%, and +20% pressure.
  • Curve equation based on retention time vs. pressure is generated.
  • Locked method title bar indicates method is locked and which compound was used.
  • When a locked method is run, pressure is restored to locked pressure even if changed via GC keypad or Instrument Control.
pp. 23
Operating Tips

Practical tips for operating the GC/MSD system, including data backup, tune file selection, maintenance scheduling, pumpdown wait time, column flow rates, injection techniques, and SIM/Scan mode selection.

  • After pumpdown, wait at least 2 hours for thermal equilibrium before tuning or acquiring data.
  • Optimum sensitivity generally occurs at column flow rates of 1.2 ml/minute or less.
  • For injections greater than 1 µL, use pulsed splitless mode and increase initial oven temperature 10–20°C.
  • For splitless injections, a pulse pressure of twice the initial inlet pressure is typical.
  • Use Constant Flow mode for most efficient separation and constant sensitivity.
  • For SIM, use exact mass from Tabulation report, not nominal mass.
  • Use low resolution mode for SIM unless determining ratios of masses one amu apart.
pp. 24-25
Troubleshooting Tips

Lists common symptoms and their probable causes, including status messages, peak shape problems, sensitivity issues, retention time drift, high background, and vacuum problems. Refers to MSD Hardware Manual and GC Maintenance and Troubleshooting Manual for more detail.

  • Status flashing 'Server not found! Check LAN connection' is normal initially; if persistent, check for power failure or bad connection.
  • No peaks: check sample concentration, syringe installation, vial, injection mode.
  • Tailing peaks: active sites, injection too large, injection port too cool, column flow too low, interface or ion source too cool.
  • Poor sensitivity: incorrect tuning, tune file mismatch, incorrect temperatures, leaks, dirty ion source, air leak, filament, mass filter polarity.
  • Foreline or vacuum manifold pressure too high: excessive column flow, air leak, diffusion pump fluid level/contamination, foreline pump oil level/contamination, constricted foreline hose.
pp. 26
Maintenance Schedule

Provides a schedule of daily, weekly, monthly, quarterly, semi-annual, and annual maintenance tasks, plus as-needed tasks. Includes safety warnings for maintenance.

  • Daily: check/replace septum, check tightness of injection port liners and column nuts.
  • Weekly: check foreline pump fluid level, change injection port liners and O-rings.
  • Monthly: clean split/splitless inlet vent line trap, check for leaks.
  • Every three months: replace gas cylinders when below 500 psig.
  • Every six months: replace foreline pump fluid, refill calibration vial if needed.
  • Every year: check/replace diffusion pump fluid, recondition/replace traps and chemical filters.
  • As needed: tune MSD, clean ion source, replace carrier gas trap, replace worn parts, replace column, lubricate seals.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Carrier gas purity must be at least 99.999% with recommended traps. p.17 ↗ Use high-purity carrier gas and appropriate traps to prevent contamination and ensure proper MSD operation. 5973N MSD with 6890 GC
If hydrogen is used as carrier gas, carrier gas flow must be off and the front sideplate thumbscrew loosely fastened before pump down. p.17 ↗ Prevents hydrogen accumulation and explosion risk; follow Hydrogen Carrier Gas Safety Guide (5955-5398). 5973N MSD with hydrogen carrier gas
After pump down, wait two hours for the MSD to reach thermal equilibrium before tuning or acquiring data. p.17 ↗ Data acquired before thermal equilibrium may not be reproducible. 5973N MSD
Turbo pump should reach 80% speed within 10 to 15 minutes and eventually 95%. p.17 ↗ If pump speed does not reach these levels, check for air leaks or vacuum problems. 5973N MSD with turbo pump
Optimum sensitivity generally occurs at column flow rates of 1.2 ml/minute or less. p.23 ↗ Use column flow rates at or below 1.2 ml/min for best sensitivity. 5973N MSD with 6890 GC
When injecting volumes greater than one microliter, use pulsed splitless mode and increase the initial oven temperature 10–20°C. p.23 ↗ Improves sample transfer and peak shape for large injections. 5973N MSD with 6890 GC
For splitless injections, a pulse pressure of twice the initial inlet pressure is typical. p.23 ↗ Provides more quantitative sample transfer onto the column. 5973N MSD with 6890 GC
When choosing masses for SIM, use the exact mass printed in the Tabulation report, not the nominal mass annotated on the spectrum display. p.23 ↗ Provides more accurate SIM data. 5973N MSD with ChemStation
For SIM analysis, use low resolution mode unless determining ratios of masses one amu apart. p.23 ↗ Low resolution provides maximum sensitivity and repeatability. 5973N MSD with ChemStation
Set GC/MSD interface heater and GC oven to ambient (25°C) before venting the MSD. p.16 ↗ Ensures safe venting and prevents damage to components. 5973N MSD
Turn vent valve knob counterclockwise only 3/4 turns or until hissing sound is heard. p.16 ↗ Prevents O-ring from falling out of its groove; retighten knob before pumping down. 5973N MSD
On a CI MSD, the Gas Off light must be on when the MSD is venting. p.16 ↗ Ensures reagent gas flow is off and isolation valve is closed. 5973N MSD CI variant
When a locked method is run, the pressure is restored to the locked pressure value even if changes were made using the GC keypad or Instrument Control. p.22 ↗ Retention time locking overrides manual pressure changes; ensure method is unlocked if pressure changes are needed. 5973N MSD with ChemStation RTLock
RTL calibration collects data at five inlet pressures: -20%, -10%, nominal, +10%, and +20% of current method setpoint. p.22 ↗ Generates pressure/retention time curve for locking; five runs are made automatically. 5973N MSD with ChemStation RTLock
Replace gas cylinders when pressure is below 500 psig. p.26 ↗ Prevents contamination and ensures consistent gas supply. 5973N MSD with 6890 GC
Check foreline pump fluid level every week and replace fluid every six months. p.26 ↗ Maintains proper vacuum and prevents pump damage. 5973N MSD
Clean the split/splitless inlet vent line trap every month. p.26 ↗ Prevents contamination and maintains proper inlet performance. 6890 GC with split/splitless inlet
For manual injection, press the Prep Run key on the GC keypad to cancel gas saver flow, bring inlet flow to setpoint, and close purge valve (for splitless injection only). p.19 ↗ Prepares the GC for manual injection and ensures proper sample transfer. 5973N MSD with 6890 GC manual injection
Do not use the Start button on the GC to start a run when using the autosampler. p.19 ↗ Use the ChemStation software to start runs with the autosampler to avoid inconsistent results. 5973N MSD with 6890 GC and ALS
Do not put the ChemStation into the Top view when the MSD is vented; doing so will turn on the interface heater. p.16 ↗ Prevents overheating and potential damage during venting. 5973N MSD

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Server not found! Check LAN connection p.24 ↗ Normal when MSD is initially turned on; if flashing continues after pumpdown initiated: temporary power failure interrupted communications, or bad connection between MSD and ChemStation and/or hub. Check power and LAN connections; if persistent, refer to MSD Hardware Manual or online help. Troubleshooting Tips
No peaks p.24 ↗ Incorrect sample concentration; no analytes present; syringe missing or not installed correctly (ALS only); empty sample vial; injection in split mode instead of splitless mode. Check sample concentration, syringe installation, vial, and injection mode. Troubleshooting Tips
Tailing peaks p.24 ↗ Active sites in sample path; injection too large; injection port too cool; column flow too low; GC/MSD interface or ion source too cool. Check for active sites, reduce injection volume, increase temperatures, increase column flow. Troubleshooting Tips
Peaks with flat tops p.24 ↗ Solvent delay time too short; display scale is wrong; injection too large; electron multiplier voltage too high. Adjust solvent delay, display scale, injection volume, or EM voltage. Troubleshooting Tips
Peaks with split tops p.24 ↗ Bad injection technique; injection too large. Improve injection technique or reduce injection volume. Troubleshooting Tips
Rising baseline p.24 ↗ Column bleed; other contamination. Condition column or check for contamination. Troubleshooting Tips
Retention time (RT) drift p.24 ↗ Column has been shortened (shorter RT); old column (shorter RT); active sites in sample path (longer RT); reduced column flow (longer RT); injection port leak (longer RT); initial oven temperature changed (up = shorter RT, down = longer RT). Check column, flow, leaks, and oven temperature program. Troubleshooting Tips
Poor sensitivity p.24 ↗ Incorrect tuning; tune file does not match type of analysis; incorrect temperatures; incorrect sample concentration; leaking injection port; incorrect split ratio; purge off time in splitless mode too short; excessive pressure in the MSD; dirty ion source; air leak; detector is not working correctly; poor filament operation; incorrect mass filter polarity. Check tuning, temperatures, concentrations, leaks, split ratio, purge time, pressure, ion source cleanliness, detector, filament, and mass filter polarity. Troubleshooting Tips
Poor repeatability p.25 ↗ Dirty syringe needle; leaking injection port; mismatched injection port liner and injection size; loose column connections; variations in pressure, column flow, and temperature; dirty ion source; loose connections in the analyzer; ground loop. Clean syringe, check for leaks, match liner to injection size, tighten connections, stabilize pressure/flow/temperature, clean ion source, check for ground loops. Troubleshooting Tips
Inconsistent peakwidths p.25 ↗ Incorrect tuning; no PFTBA in calibration vial; calibration valve failure; dirty ion source; worn out electron multiplier; MSD has not had enough time to reach thermal equilibrium; large variations in the temperature of the lab. Check tuning, calibration vial, calibration valve, ion source, electron multiplier, allow thermal equilibrium, stabilize lab temperature. Troubleshooting Tips
High background in mass spectra p.25 ↗ Air leak; foreline or vacuum manifold pressure too high; other contamination. Check for air leaks, vacuum pressure, and contamination. Troubleshooting Tips
Ions at m/z 18, 28, 32, and 44 p.25 ↗ Detector vented recently (residual air and water); air leak. Allow system to pump down longer; check for air leaks. Troubleshooting Tips
Isotopes missing or isotope ratios incorrect p.25 ↗ Incorrect tuning; dirty ion source; high background; electron multiplier voltage too high; repeller voltage too high; high scan speed (Scan mode); low dwell time (SIM mode); peaks too wide or too narrow; repeller and ion focus leads have been reversed. Check tuning, clean ion source, reduce background, adjust EM and repeller voltages, scan speed, dwell time, peak width, and verify lead connections. Troubleshooting Tips
Foreline or vacuum manifold pressure too high p.25 ↗ Excessive column flow; air leak; diffusion pump fluid level too low; diffusion pump fluid is contaminated; foreline pump oil level too low; foreline pump oil is contaminated; constricted foreline hose (this would cause the vacuum manifold pressure to be too high but the foreline pressure to be too low). Check column flow, air leaks, diffusion pump fluid level and condition, foreline pump oil level and condition, and foreline hose for constriction. Troubleshooting Tips