Induztr
Engineering Digest Document Viewer Related Models

Fast Hydrocarbon and Sulfur Simulated Distillation Using the

Technical Reference Summary

Covered Models & Document Scope

Application note describing a fast multielement simulated distillation (SimDis) method for hydrocarbon and sulfur boiling point distribution using the Agilent Low Thermal Mass (LTM) system on the 7890A GC with FID and the 355 Sulfur Chemiluminescence Detector (SCD) in series configuration; no product model/size variants or ordering codes are covered.

Technical Data

Parameter Value Condition
LTM column temperature 350 °C p.2 ↗
Column DB-1 5 m × 320 mm × 1 µm p.2 ↗ LTM column module
FID temperature 350 °C p.2 ↗
H2 flow 40 mL/min p.2 ↗
Data rate 5 Hz p.2 ↗
Burner temperature 800 °C p.2 ↗ SCD
Air flow 8.3 SCCM p.2 ↗
Temperature program rate 200 °C/min p.2 ↗ example high program rate

Operating Limits

  • Run time for hydrocarbon and sulfur boiling point distribution vs conventional ASTM D2887 procedure: at least six times shorter p.1 ↗