Agilent 7820A Gas Chromatograph
Technical Reference Summary
Covered Models & Document Scope
Agilent 7820A Gas Chromatograph System; configurations with EPC (electronic pneumatics control) for all inlets and detectors, and EPR (electronic pneumatics regulation) for select inlets and detectors; optional automated sample injectors and samplers; software options OpenLAB CDS Workstation/Workstation Plus, OpenLAB CDS VL Workstation/Workstation Plus, OpenLAB CDS ChemStation Edition (C.01.05 or higher), EZChrom Edition (A.04.05 or higher), OpenLAB CDS ChemStation VL, EZChrom VL, OpenLAB CDS EZChrom Compact, DA Express (China, Russia only)
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Operating temperature | 8 °C above ambient to 425 °C °C p.2 ↗ | |
| Mains frequency | 47.5~63 Hz p.2 ↗ | |
| Maximum operating temperature (inlet) | 400 °C p.3 ↗ | Purged Packed (EPC) |
| Maximum split ratio | 250:1 p.3 ↗ | |
| Minimum detection level (MDL) | < 3 pg carbon/s as tridecane p.3 ↗ | |
| Linear dynamic range | > 107 p.3 ↗ | with N2 carrier |
| Maximum data acquisition rate | 100 Hz p.3 ↗ | |
| NPD selectivity | 25,000 to 1 gN/gC, 75,000 to 1 gP/gC p.3 ↗ | NPD (nitrogen phosphorous detector) |
| NPD dynamic range | > 104 N, > 104 P p.3 ↗ | NPD (nitrogen phosphorous detector) |
Operating Limits
- Operating temperature range: 8 °C above ambient to 425 °C p.2 ↗
- Mains frequency range: 47.5~63 Hz p.2 ↗
- Maximum inlet operating temperature (Purged Packed, EPC): 400 °C p.3 ↗
- Maximum split ratio: 250:1 p.3 ↗
- Flow adjustable range: 0 to 200 mL/min N2 p.3 ↗