Induztr
Engineering Digest Document Viewer Related Models

Automated Scouting of Stationary and Mobile Phases Using the Agilent 1290 Infinity II

Technical Reference Summary

Covered Models & Document Scope

Agilent 1290 Infinity II Method Development Solution (Technical Overview) covering automated scouting of stationary and mobile phases; modules: 1290 Infinity II Flexible Pump (G7104A), 1290 Infinity II Multisampler (G7167B), 1290 Infinity II Multicolumn Thermostat (G7116B) with valve drive option #058 and Quick-Change column selection valve (G4239C, 1,300 bar); software OpenLAB CDS ChemStation Edition rev. C.01.07 with Method Scouting Wizard rev. A.02.05; ZORBAX RRHD columns 2.1 × 50 mm, 1.8 µm (SB-C18, Eclipse Plus C18, Eclipse Plus C8, Extend C18, SB-CN, Bonus RP, Eclipse Plus Phenyl-Hexyl)

Technical Data

Parameter Value Condition
Column selection valve pressure rating 1,300 bar p.2 ↗ G4239C
Number of columns switchable up to eight columns p.1 ↗ 1290 Infinity II Multicolumn Thermostat
Number of solvents connectable up to 26 solvents p.1 ↗ Flexible Pump with two solvent-selection valves
Column dimensions 2.1 × 50 mm p.2 ↗ ZORBAX RRHD
Column particle size 1.8 µm p.2 ↗ ZORBAX RRHD
Detection wavelength 254/10 nm p.8 ↗ reference
Software revision C.01.07 p.2 ↗ OpenLAB CDS ChemStation Edition
Minimum driver requirement A.02.11 SP1 p.2 ↗ OpenLAB CDS ChemStation Edition
Method Scouting Wizard revision A.02.05 p.2 ↗

Operating Limits

  • Maximum pressure of Quick-Change column selection valve: 1,300 bar p.2 ↗
  • Maximum number of columns in Multicolumn Thermostat: 8 columns p.1 ↗
  • Maximum number of solvents with two solvent-selection valves: 26 solvents p.1 ↗

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
Agilent ZORBAX RRHD SB-C18, 2.1 × 50 mm, 1.8 µm p.2 ↗ 857700-902 column
Agilent ZORBAX RRHD Eclipse Plus C18, 2.1 × 50 mm, 1.8 µm p.2 ↗ 959757-902 column
Agilent ZORBAX RRHD Eclipse Plus C8, 2.1 × 50 mm, 1.8 µm p.2 ↗ 959757-906 column
Agilent ZORBAX RRHD Extend C18, 2.1 × 50 mm, 1.8 µm p.2 ↗ 757700-902 column
Agilent ZORBAX RRHD SB-CN, 2.1 × 50 mm, 1.8 µm p.2 ↗ 857700-305 column
Agilent ZORBAX RRHD Bonus RP, 2.1 × 50 mm, 1.8 µm p.2 ↗ 857768-901 column
Agilent ZORBAX RRHD Eclipse Plus Phenyl-Hexyl, 2.1 × 50 mm, 1.8 µm p.2 ↗ 959757-912 column