An Optimal Method for the Analysis of Pesticides in a Variety of Matrices
Technical Reference Summary
Covered Models & Document Scope
Application note covering an analytical method for 195 pesticides in eight matrices (extra virgin olive oil, black loose leaf tea, jasmine rice, cucumber, organic honey, navel orange, yellow onion, baby spinach) plus ACN, using Agilent QuEChERS kits and GC/MS/MS MRM transitions; no instrument model variants or ordering codes are defined.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Injection mode | Hot-splitless p.3 ↗ | MMI |
| Injection volume | 1 µL p.3 ↗ | |
| Inlet temperature | 280 °C p.3 ↗ | |
| Carrier gas | He, constant flow 1.00 mL/min p.3 ↗ | |
| Source temperature | 300 °C p.3 ↗ | |
| Quad temperatures | 150 °C p.3 ↗ | |
| Column | two HP-5ms UI columns (15 m × ...), p/n 19091S-431 UI p.3 ↗ | coupled |
| QuEChERS EN salts part number | 5982-5650 p.2 ↗ | |
| QuEChERS dSPE general fruit/vegetable part number | 5982-5056 p.2 ↗ | |
| QuEChERS dSPE fatty part number | 5982-5156 p.2 ↗ | |
| Sample amount, extra virgin olive oil | 3 g oil/7 mL water p.2 ↗ | High oil matrix |
| Sample amount, black loose leaf tea | 3 g tea/7 mL water p.2 ↗ | Difficult matrix |
| Sample amount, jasmine rice | 3 g rice/7 mL water p.2 ↗ | High starch matrix |
| Sample amount, cucumber | 10 g p.2 ↗ | High water matrix |
| Sample amount, organic honey | 5 g honey/5 mL water p.2 ↗ | High sugar matrix |
| Sample amount, navel orange | 10 g p.2 ↗ | High acid matrix |
| Centrifugation speed | 5,000 rpm p.3 ↗ | |
| Vortex time | 2 minutes p.3 ↗ | |
| Number of target compounds | 195 p.1 ↗ | |
| Calibration curve R² criterion | ≥ 0.990 p.1 ↗ | 90% of target compounds |
| Repeatability %RSD limit | ≤ 30 % p.1 ↗ | all analyzed pesticides |
| Limit of quantitation (LOQ) criterion | ≤ 1.5 pg/µL p.1 ↗ | 90% of analyzed pesticides |
Operating Limits
- Maximum %RSD of repeated measurements: ≤ 30% p.1 ↗
- Minimum calibration curve R² for 90% of target compounds: ≥ 0.990 p.1 ↗
- Maximum LOQ for 90% of analyzed pesticides: ≤ 1.5 pg/µL p.1 ↗