Agilent AdvanceBio Workflows for Spent Media Analysis
Technical Reference Summary
Covered Models & Document Scope
AdvanceBio Amino Acid Analysis columns and kit (derivatized LC/UV workflow) and AdvanceBio MS Spent Media columns (underivatized LC/MS workflow); MS Spent Media columns in 120 Å, 2.1 x 5 mm, 2.1 x 100 mm and 2.1 x 150 mm
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Particle size | 2.7 µm p.4 ↗ | Poroshell particles, both column families |
| Pressure tolerance | up to 600 bar p.6 ↗ | AdvanceBio MS Spent Media, PEEK-lined stainless steel |
| Column temperature | 30 °C p.7 ↗ | AdvanceBio MS Spent Media 2.1 x 100 mm, LC/MS method |
| Mobile phase A | 10% 200 mM ammonium formate in water pH 3, 90% water p.7 ↗ | AdvanceBio MS Spent Media, low pH positive ion mode |
| Mobile phase B | 10% 200 mM ammonium formate in water pH 3, 90% acetonitrile p.7 ↗ | AdvanceBio MS Spent Media, low pH positive ion mode |
| Final salt concentration | 20 mM p.7 ↗ | AdvanceBio MS Spent Media mobile phase |
| Resolution leucine/isoleucine | 3.0 p.7 ↗ | AdvanceBio MS Spent Media 2.1 x 100 mm, TOF positive ion mode |
| Resolution requirement (European Pharmacopoeia) | >1.5 p.7 ↗ | European Pharmacopoeia 9.0 (2.2.56) |
Operating Limits
- Max pressure tolerance: 600 bar p.6 ↗
- Recommended column temperature: 30 °C p.7 ↗
Ordering Data
| Ordering Code | Part No. | Variant / Code | Applies To | Note |
|---|---|---|---|---|
| Standards and Reagents kit p.7 ↗ | 5190-9426 | AdvanceBio AAA kit contents | ||
| Buffer, borate, 100 mL p.7 ↗ | 5061-3339 | Kit content, can be ordered separately | ||
| AdvanceBio MS Spent Media 120 Å, 2.1 x 5 mm p.7 ↗ | 679775-901 | |||
| AdvanceBio MS Spent Media 120 Å, 2.1 x 100 mm p.7 ↗ | 675775-901 | |||
| AdvanceBio MS Spent Media 120 Å, 2.1 x 150 mm p.7 ↗ | 673775-901 |