Agilent 6400 Series Triple Quad LC/MS Concepts Guide
Technical Reference Summary
Covered Models & Document Scope
Concepts guide for the Agilent 6400 Series Triple Quad LC/MS, covering models 6420, 6460, 6470, 6495A and 6495B, including ion source options ESI, APCI, AJS, APPI and MMI, and MassHunter Workstation Data Acquisition software.
Section-by-Section Technical Breakdown
The Triple Quad LC/MS can be configured for normal-flow, microflow or nanoflow LC/MS. Supported ion sources include ESI, APCI, APPI and MMI; the 6460, 6470 and 6495 support the AJS source, and the 6495 adds iFunnel technology.
- Normal flow: binary or quaternary pump, well-plate sampler/autosampler/CTC PAL, sources ESI, APCI, APPI, MMI
- Microflow: capillary pump, micro well-plate sampler or CTC PAL micro-plate autosampler, sources ESI, APCI or MMI
- Nanoflow: nanoflow pump, capillary pump and micro well-plate sampler for narrow peak dispersion and lower dead volumes
- 6460, 6470 and 6495 can use the AJS source with super-heated sheath gas
- 6495 uses iFunnel technology: dual offset ion funnel, AJS source, shorter desolvation assembly with hexabore capillary
- 6495 and 6470 have new ion optics, CATT2 curved collision cell and a new detector
Dynamic MRM uses a single continuous time segment with up to 4000 transitions. Transitions are automatically grouped into MRM tables of up to 200 transitions based on retention time and delta retention time, allowing longer dwell times around compound elution.
- Up to 4000 transitions in the Scan Segments table
- Transitions grouped into MRM tables of up to 200 transitions each
- Up to 1000+ MRM tables can be created
- Ret Time and Delta Ret Time entered in minutes
- Each transition acquired from Ret Time - 1/2*(Delta Ret Time) to Ret Time + 1/2*(Delta Ret Time)
- DSP controls the run and switches MRM tables with minimal delay and no data loss
Triggered Dynamic MRM acquires confirmatory secondary MRMs when primary trigger MRM abundances exceed set thresholds. Up to two primary MRMs per compound can be designated as triggers, and secondary acquisition is controlled by Trigger Entrance Delay, Trigger Delay and Trigger Window.
- Up to two Trigger MRMs per compound
- Secondary transitions acquired only when trigger abundances meet or exceed thresholds
- Trigger Entrance Delay skips scans after threshold is met
- Trigger Delay sets scans between secondary acquisitions
- Trigger Window defaults to 0, meaning the full Peak Retention Time window
- Matrix-matched standards are preferable for setting trigger thresholds
The guide describes noise-reduction design features and reports linear dynamic ranges for verapamil: six orders of magnitude for 6470 and 6495, and five orders of magnitude for 6460 and 6420.
- 6470 and 6495: six orders of magnitude linear dynamic range
- 6460 and 6420: five orders of magnitude linear dynamic range
- Curved high-pressure collision cell eliminates crosstalk even at very low dwell times
- Dual high-energy dynode detector permits high gain with rapid polarity switching
- Orthogonal spray sources minimize solvent and matrix noise
The guide describes AJS, ESI, APCI and multimode ion sources. The AJS source uses superheated nitrogen sheath gas for thermal gradient focusing; the 6495 uses a hexabore capillary sampling up to 10 times more ion-rich gas.
- AJS parameters: superheated nitrogen sheath gas temperature and flow rate, and nozzle voltage
- 6495 hexabore capillary has 6 capillary inlets and samples up to 10X more ion-rich gas
- 6460 and 6420 use a resistive capillary allowing virtually instantaneous polarity switching
- APCI operates at higher temperatures and does not generate multiply charged ions
- G1978B Multimode Source can operate in ESI and APCI simultaneously
- Simultaneous ESI and APCI operation can lose sensitivity up to a factor of two for some compounds
The collision cell is a high-pressure hexapole assembly with linear acceleration to optimize MS/MS fragmentation and eliminate crosstalk. The 6470 and 6495 use a hexapole field axial focusing curved collision cell with a tapered entrance.
- High-pressure hexapole collision cell with linear acceleration
- Small diameter and high frequency hexapole design
- Curved collision cell reduces ionizer-generated noise
- Tapered cell structure increases ion acceptance at the entrance
- Eliminates crosstalk even at very low dwell times
The 6495 uses two rough pumps and a three-stage turbo pump for the first five vacuum stages. The second turbo pump is added to the last vacuum stage to handle gas load from the curved collision cell.
- 6495: two rough pumps and a three-stage turbo pump
- Two roughing pumps decrease pressure in the dual-stage ion funnel
- Second turbo pump added to the last vacuum stage
- Second turbo pump helps pump gas load from the hexapole field axial focusing curved collision cell
Checktune verifies peak width and mass axis before acquisition and takes approximately 3 to 5 minutes per polarity. Autotune is needed after preventative maintenance or if Checktune fails; it takes approximately 15 to 20 minutes per polarity except for the 6495, which takes approximately 25 minutes per ionization mode.
- Checktune duration: approximately 3 to 5 minutes per polarity
- Autotune duration: approximately 15 to 20 minutes per polarity for all models except 6495
- 6495 Autotune duration: approximately 25 minutes per ionization mode
- Tuning mix delivered automatically by the calibrant delivery system (CDS)
- Mark Tune from factory default only if regular Autotune fails or the tune file is corrupted
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| The 6470 has a mass range of 5 to 3000 m/z. p.22 ↗ | Suitable for analytes requiring high m/z coverage up to 3000. | 6470 |
| The 6495A has a mass range of 5 to 2250 m/z; the 6495B extends the mass range to 3000 amu. p.23 ↗ | Select 6495B when m/z above 2250 is required. | 6495A, 6495B |
| Linear dynamic range for verapamil is six orders of magnitude for 6470 and 6495, and five orders of magnitude for 6460 and 6420. p.49 ↗ | Use 6470 or 6495 for applications requiring the widest quantitation range. | 6470, 6495, 6460, 6420 |
| Dynamic MRM supports up to 4000 transitions and automatically groups them into MRM tables of up to 200 transitions each. p.33 ↗ | Large MRM methods can be run with longer dwell times around compound elution. | Dynamic MRM |
| Triggered Dynamic MRM allows up to two Trigger MRMs per compound. p.35 ↗ | Confirmatory secondary transitions can be acquired only when primary abundances exceed thresholds. | Triggered Dynamic MRM |
| The Trigger Window defaults to 0, meaning it equals the Peak Retention Time window. p.37 ↗ | Set a narrower Trigger Window to restrict threshold monitoring to a sub-window. | Triggered Dynamic MRM |
| The 6495 hexabore capillary has 6 capillary inlets and samples up to 10X more ion-rich gas from the source. p.27 ↗ | Provides higher ion transmission and sensitivity for the 6495. | 6495 |
| The first ion funnel operates between 7 and 14 Torr; the second low-pressure ion funnel operates between 1 and 3 Torr. p.28 ↗ | Dual-stage ion funnel removes gas and neutral noise while transmitting ions. | 6495 iFunnel |
| The 6495 uses two rough pumps and a three-stage turbo pump for the first five vacuum stages. p.62 ↗ | Requires additional roughing pump capacity and maintenance compared with other models. | 6495 |
| Checktune takes approximately 3 to 5 minutes per polarity. p.69 ↗ | Routine mass axis and peak width verification can be performed before each acquisition. | All models |
| Autotune takes approximately 15 to 20 minutes per polarity for all models except the 6495, which takes approximately 25 minutes per ionization mode. p.69 ↗ | Plan longer tuning time for the 6495. | All models |
| The 6495B Gate Valve Technology allows capillary replacement without venting the instrument. p.26 ↗ | Higher uptime and productivity; a Capillary Removal Tool is included. | 6495B |
| The 6460 and 6420 use a resistive capillary that allows virtually instantaneous polarity switching. p.53 ↗ | Fast polarity switching is supported on these models. | 6460, 6420 |
| Simultaneous ESI and APCI operation in the multimode source can lose sensitivity up to a factor of two for some compounds. p.54 ↗ | For mixed-mode operation, verify sensitivity per compound; single-mode operation shows no loss for compounds studied. | G1978B Multimode Source |
| The collision region between capillary exit and skimmer on the single quad LC/MS is at about 2 Torr, about three orders of magnitude below atmosphere. p.20 ↗ | Collisionally induced dissociation can be performed in this low-pressure region. | Single quad LC/MS |
| The 6495B includes Gate Valve Technology with a safety lock to prevent accidental opening or closing. p.26 ↗ | Capillary maintenance can be performed without venting, reducing downtime. | 6495B |
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Mass range | 5 to 3000 m/z p.22 ↗ | 6470 |
| Mass range | 5 to 2250 m/z p.23 ↗ | 6495A |
| Mass range | 3000 amu p.26 ↗ | 6495B |
| Linear dynamic range | six orders of magnitude p.49 ↗ | 6470 and 6495, verapamil |
| Linear dynamic range | five orders of magnitude p.49 ↗ | 6460 and 6420, verapamil |
| Limit of detection and quantitation | Femtogram-level p.9 ↗ | 6460 |
| Limit of detection and quantitation | Attogram-level p.9 ↗ | 6470 and 6495 |
| First ion funnel pressure | 7 to 14 Torr p.28 ↗ | 6495 iFunnel |
| Second ion funnel pressure | 1 to 3 Torr p.28 ↗ | 6495 iFunnel |
| Collision region pressure | about 2 Torr p.20 ↗ | single quad LC/MS, capillary exit to skimmer |
| Maximum Dynamic MRM transitions | 4000 transitions p.33 ↗ | Dynamic MRM |
| Maximum transitions per MRM table | 200 transitions p.33 ↗ | Dynamic MRM |
| Maximum number of MRM tables | 1000+ tables p.33 ↗ | Dynamic MRM |
| Maximum Trigger MRMs per compound | two p.35 ↗ | Triggered Dynamic MRM |
| Checktune duration | 3 to 5 minutes p.69 ↗ | per polarity |
| Autotune duration | 15 to 20 minutes p.69 ↗ | all models except 6495, per polarity |
| Autotune duration | 25 minutes p.69 ↗ | 6495, per ionization mode |