Introduces the 30-color immunophenotyping panel purpose and context
Procedurep. 3
Step-by-step cell pellet centrifugation and staining protocol
pp. 4-5
Presents results and discussion of panel performance
p. 6
Gating strategy for defining cell subpopulations
p. 7
Frequencies of main subpopulations across samples
p. 8
Summarizes findings and panel utility
p. 3table, 30 rows
Table 1. Antibodies used in 30-color immunophenotyping panel on the Agilent NovoCyte Penteon Flow Cytometer, ordered by channels. The laser and filter used for detection for each antibody can be found in the table, as well as concentration and product details for each antibody.
Laser
Filter
Specificity
Fluorochrome
Clone
Vendor
Part Number
Dilution
Amount Added to 3 × 106 Cells (50 μL)
349
445/45
Live/Dead Fixable Blue
BUV395
Invitrogen
L23105
1/1000
0.05
525/45
MHC-II
BUV496
M5/114.15.2
Becton Dickinson
750281
1/800
0.0625
586/20
CD3
BUV563
17A2
Becton Dickinson
741319
1/400
0.125
615/20
gdT
BUV615
GL3
Becton Dickinson
751183
1/200
0.25
667/30
CD88
BUV661
20/70
Becton Dickinson
750080
1/400
0.125
725/40
CD43
BUV737
S7
Becton Dickinson
612840
1/200
0.25
780/60
NKp46
BUV805
29A1.4
Becton Dickinson
742066
1/100
0.5
405
445/45
XCR1
BV421
ZET
BioLegend
148216
1/400
0.125
525/45
CD103
BV480
M290
Becton Dickinson
566118
1/200
0.25
586/20
TCRb
BV570
H57-597
BioLegend
109231
1/200
0.25
615/20
Siglec H
sb600
eBio440c
Thermo Fisher Scientific
63-0333-82
1/200
0.25
667/30
CD11b
BV650
M1/70
BioLegend
101259
1/400
0.125
725/40
CD8b
sb702
53-6.7
Thermo Fisher Scientific
67-0081-82
1/400
0.125
780/60
CD69
BV786
H1.2F3
Becton Dickinson
564683
1/200
0.25
488
525/45
SiglecF
BB515
E50-2440
Becton Dickinson
564514
1/400
0.125
586/20
CD45
Spark 574
30F11
BioLegend
103184
1/200
0.25
615/20
CD4
BB630-P
GK1.5
Becton Dickinson
Prototype
1/400
0.125
667/30
CD90.2
PerCP
30H12
BioLegend
105322
1/200
0.25
695/40
Ly6G
BB700
1A8
Becton Dickinson
566435
1/800
0.0625
725/40
CD62L
BB755
MEL-14
Becton Dickinson
Prototype
1/400
0.125
561
586/20
B220
SYG570
RA36B2
BioLegend
103286
1/400
0.125
615/20
CD172a
AF594
P84
BioLegend
144020
1/400
0.125
667/30
CD25
PE Cy5
PC61
BioLegend
102010
1/200
0.25
695/40
CD11c
PE Cy5.5
N418
Thermo Fisher Scientific
35-0114-82
1/400
0.125
725/40
CD19
PE-AF700
1D3
AAT-bioquest
ATB101941O0
1/400
0.125
780/60
NK1.1
PE-Cy7
PK136
BioLegend
108714
1/100
0.5
637
667/30
CD64
AF647
X54-5/7.1
BioLegend
139322
1/200
0.25
695/40
IgD
SPARK 685
11-26c.2a
BioLegend
405750
1/1000
0.05
725/40
CD44
APC-R700
IM7
Becton Dickinson
565480
1/400
0.125
780/60
Ly6C
APC-Cy7
HK1.4
BioLegend
128026
1/800
0.0625
Engineering Facts
12 facts · every claim linked to a page
Selection
The instrument must be specified with all five laser wavelengths to run this 30-color panel.
Laser and detector count defines whether the full 30-color panel can be acquired in a single tube.
Source
“the Novocyte Penteon is equipped with five lasers (UV, 349 nm; violet, 405 nm; blue, 488 nm; yellow-green, 561 nm; and red, 637 nm) and 30 fluorescence detectors”
Selection
Panel design must account for spectral overlap and the specific instrument configuration, not just marker expression.
Fluorophore choice is constrained by the detector/filter layout, so the panel must be validated against the actual instrument configuration.
Source
“designed through diligent fluorophore selection considering expression markers, fluorescence intensity, spectral overlap, and Agilent NovoCyte Penteon configuration”
Operation
Tissue dissociation uses a two-pass enzymatic protocol with a defined 30-minute incubation at 37 °C.
Reproducible sample prep depends on holding the specified temperature and timing between dissociation passes.
Source
“Wait 30 minutes at 37 °C. Run the "m_lung-01_02" program again on OctoGentleMacs (Miltenyi).”
Operation
EDTA is added after dissociation and held for 2 minutes before filtering.
The EDTA quench step must be timed and dosed exactly to stop enzymatic digestion consistently.
Source
“Add 30 µL of EDTA (0.5 M) and wait 2 minutes.”
Operation
All centrifugation steps are specified at 400 g and 4 °C for 7 minutes.
The centrifuge must reliably hold 400 g at 4 °C to preserve cell viability and pellet recovery.
Source
“Centrifuge for 7 minutes at 4 °C, 400 g.”
Operation
Cells are stained for 30 minutes at 37 °C in the dark, which requires a temperature-controlled dark incubation capability.
Staining temperature and light protection directly affect panel performance and must be controllable.
Source
“Incubate for 30 minutes at 37 °C in the dark.”
Operation
Viability dye is used at a 1/1,000 dilution in PBS with a 15-minute room-temperature dark incubation.
Dilution accuracy and ambient temperature control are needed for consistent live/dead discrimination.
Source
“Resuspend the cells in 100 µL of Live Dead Fixable blue (diluate at 1/1,000 in PBS) for 15 minutes at room temperature (RT) in the dark.”
Performance
The panel is designed to use the maximum detector count available on a single cytometer.
Confirms the configuration ceiling and that all 30 detectors are utilized, so no spare channels are assumed.
Source
“Design of a 30-color immunophenotyping panel maximizes throughput on a 30-detector instrument—to date the maximum number of detectors on a single flow cytometer.”
Operation
The antibody cocktail volume is fixed at 50 µL per 3 million cells.
Staining volume and cell number must be matched to keep the panel reproducible.
Source
“To generate the panel, 50 µL of the antibody cocktail was added to 3 million cells.”
Performance
Absolute counts in events/µL are computed by the software during acquisition.
Volumetric absolute counting must be supported by the acquisition software if concentration data is required.
Source
“The absolute count (on event/µL) of each sample was also determined by NovoExpress Software during acquisition.”
Operation
A single gating strategy is applied across all three organs to allow direct comparison.
Consistent gating across samples is required for valid cross-organ comparison in analysis.
Source
“The same gating strategy was used across all three organs to easily compare the data.”
Selection
The system enables multi-parameter acquisition from a single tube, conserving sample material.
Single-tube acquisition reduces sample consumption, which matters when biological material is limited.
Source
“With five lasers and 30 detectors, it is possible to extract more information from a single tube, saving precious biological material.”