Accurate Mass Library for Natural Products Based on Compounds Identified in Hemp Oil
Technical Reference Summary
Covered Models & Document Scope
Application note covering an accurate-mass EI Natural Products spectral library for GC/Q-TOF MS built from comprehensive GC×GC data of hemp CBD oil; describes GC and MS conditions for 2D GC×GC and 1D GC configurations on an Agilent 7250 GC/Q-TOF with Agilent 7890B GC.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Oven temperature program | 60 °C for 5 min; 3 °C/min to 290 °C, 25 min hold °C p.2 ↗ | 2D GC×GC |
| Modulation period | 6 s p.2 ↗ | 2D GC×GC |
| Cold jet flow | 13 L/min p.2 ↗ | 2D GC×GC |
| Hot jet temperature | 300 °C p.2 ↗ | 2D GC×GC |
| Hot jet duration | 320 ms p.2 ↗ | 2D GC×GC |
| Transfer line temperature | 280 °C p.2 ↗ | |
| Quadrupole temperature | 150 °C p.2 ↗ | |
| Source temperature | 200 °C p.2 ↗ | |
| Electron energy | 70 eV p.2 ↗ | EI mode |
| Spectral acquisition rate | 50 Hz p.2 ↗ | 2D GC×GC |
Operating Limits
- Number of chromatographic peaks in 1D configuration: 350 to 560 peaks p.2 ↗