Analysis of Cannabinoids and their Metabolites in Human Urine Using the Agilent Chem Elut S Plate by LC/MS/MS
Technical Reference Summary
Covered Models & Document Scope
Application note for the Agilent Chem Elut S 96-well plate (2 mL, square well, full skirt) used for SLE of THC, THC-OH, THC-COOH, CBD and CBN from human urine with LC/MS/MS; includes method parameters, calibration, QC and verification data.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Plate format | 96-well p.2 ↗ | Chem Elut S |
| Well volume | 2 mL p.2 ↗ | Chem Elut S |
| Well shape | square p.2 ↗ | Chem Elut S |
| Plate skirt | full skirt p.2 ↗ | Chem Elut S |
| Sorbent type | synthetic sorbent p.2 ↗ | Chem Elut S |
| LC flow rate | 0.4 mL/min p.3 ↗ | |
| Needle wash | 1:1:1:1 ACN/MeOH/IPA/water, 0.2% formic acid p.3 ↗ | |
| Ionization mode | Electrospray ionization (ESI) p.3 ↗ | |
| Nebulizer pressure | 22 psi p.3 ↗ | |
| Nozzle voltage | 0 V p.3 ↗ | both positive and negative |
| Calibration range | 2 to 1,000 ng/mL p.9 ↗ | human urine matrix |
| Linearity (R²) | >0.99 p.9 ↗ | all analytes |
| LLOQ | 2 ng/mL p.8 ↗ | |
| Urine volume per sample | 50 µL p.4 ↗ | calibration standards and QC |
| IPA pre-addition volume | 50 µL p.4 ↗ | calibration standards and QC |
| THC-COOH glucuronide spiking level | 75.75 ng/mL p.9 ↗ | corresponds to THC-COOH 50 ng/mL |
| Extraction solvent mixture | 9:1 hexane/EtOAc p.7 ↗ | elution |
| Recommended maximum elution volume | 15 % of total p.6 ↗ | not recommended to use more |
Operating Limits
- Maximum recommended elution volume: 15% of total p.6 ↗
- Calibration dynamic range: 2 to 1,000 ng/mL p.9 ↗
- LLOQ: 2 ng/mL p.8 ↗
Ordering Data
| Ordering Code | Part No. | Variant / Code | Applies To | Note |
|---|---|---|---|---|
| 400 µL (part number 5610-2004) p.3 ↗ | 5610-2004 | Sample preparation equipment/consumable referenced in Experimental section | ||
| part number 5191-4116 p.3 ↗ | 5191-4116 | Sample preparation equipment/consumable referenced in Experimental section |