Ammonia Analysis in High-Purity Hydrogen for Fuel Cell Vehicles Using an Agilent 8890 GC/8255 NCD System

Technical Reference Summary

Covered Models & Document Scope

Application note describing a GC/NCD method for trace ammonia analysis in hydrogen for fuel cell vehicles, using an Agilent 8890 GC with six-port gas sampling valve, 2 mL sample loop, NCD detector, and Agilent J&W Select Low Ammonia column (p/n CP8590). No product model variants or ordering codes are covered.

Technical Data

Parameter Value Condition
Sample loop volume 2 mL p.3 ↗
Column Select Low Ammonia (p/n CP8590) p.3 ↗
Burner temperature 900 °C p.3 ↗
Burner pressure 131 Torr p.3 ↗
Data rate 5 Hz p.3 ↗
Calibration level L1 concentration 37.1 ppb p.4 ↗
Calibration level L2 concentration 49.5 ppb p.4 ↗
Calibration level L3 concentration 79.2 ppb p.4 ↗
Calibration level L4 concentration 99 ppb p.4 ↗
Calibration level L5 concentration 198 ppb p.4 ↗
Calibration level L6 concentration 495 ppb p.4 ↗
Response factor L1 0.228 p.4 ↗
Response factor L2 0.236 p.4 ↗
Response factor L3 0.218 p.4 ↗
Response factor L4 0.215 p.4 ↗
Response factor L5 0.175 p.4 ↗
Response factor L6 0.156 p.4 ↗
Linearity range 37.1 to 495 ppb p.3 ↗
Quality control requirement 100 ppb p.2 ↗ ISO 14687-2019 / GB/T 37244-2018

Operating Limits

  • Linearity range studied: 37.1 to 495 ppb p.3 ↗
  • Quality control requirement (ISO 14687-2019, GB/T 37244-2018): 100 ppb p.2 ↗