Sensitive Quantitation of Glucagon-Like Peptide-1 Analog Semaglutide in Plasma

Technical Reference Summary

Covered Models & Document Scope

Application note covering an automated LC/MS workflow for quantitation of the GLP-1 analog semaglutide (and tirzepatide MRM transitions) in human plasma using an Agilent 1290 Infinity II bio LC, Agilent 6495D triple quadrupole LC/MS, AssayMAP Bravo with RP-S cartridges, and MassHunter software; no instrument model/size variants or ordering codes are given.

Technical Data

Parameter Value Condition
Injection volume 5 μL p.3 ↗
LLOQ 0.2 ng/mL p.1 ↗ semaglutide
Linear range upper limit 1,000 ng/mL p.1 ↗ semaglutide
Starting material volume 100 μL p.1 ↗ human plasma
Sensitivity improvement vs organic solvent precipitation five times p.1 ↗
Ion mode Positive p.4 ↗
Gas temperature 270 °C p.4 ↗
Tirzepatide precursor ion 1,204.4 m/z p.4 ↗
Tirzepatide product ion 396.2 m/z p.4 ↗
Tirzepatide collision energy 26 p.4 ↗
Tirzepatide second product ion 136.1 m/z p.4 ↗ precursor 1,029.4
Tirzepatide second collision energy 42 p.4 ↗ precursor 1,029.4
Semaglutide precursor ion 1,029.4 m/z p.4 ↗
Semaglutide product ion 1,238.1 m/z p.4 ↗
Semaglutide collision energy 30 p.4 ↗
Low QC concentration 0.5 ng/mL p.6 ↗ semaglutide
Mid QC concentration 10.00 ng/mL p.6 ↗ semaglutide
High QC concentration 800 ng/mL p.6 ↗ semaglutide
Interday % CV at LLOQ 6.7 % p.6 ↗ 0.2 ng/mL
Interday % CV at Low QC 6.3 % p.6 ↗ 0.5 ng/mL
Interday % CV at Mid QC 8.8 % p.6 ↗ 10.00 ng/mL
Interday % CV at High QC 5.5 % p.6 ↗ 800 ng/mL

Operating Limits

  • LLOQ: 0.2 ng/mL p.1 ↗
  • Upper limit of linear calibration range: 1,000 ng/mL p.1 ↗
  • Minimum starting plasma volume: 100 μL p.1 ↗