Automated Sample Prep Using Agilent PAL3 RTC for EPA 8270E SVOC Analysis by GC/TQ

Technical Reference Summary

Covered Models & Document Scope

Application note / proof-of-concept for automated sample preparation of semivolatile organic compounds (SVOCs) in water samples using a PAL3 Series II RTC system followed by GC/TQ analysis according to EPA 8270E; covers 100 SVOCs, calibration standards, method blanks and matrix-spiked QC samples.

Technical Data

Parameter Value Condition
Carrier gas Helium p.2 ↗ constant flow, 1.2 mL/min
Carrier gas flow 1.2 mL/min p.2 ↗ constant flow
Inlet liner 5190-2293, 900 μL, single taper, wool, Ultra p.2 ↗ splitless inlet liner
Ion source drawout lens diameter 9 mm p.2 ↗ 7000 series triple quadrupole MS
Ion source drawout lens part number G3870-20449 p.2 ↗ 7000 series triple quadrupole MS
Autotune algorithm etune p.2 ↗ Agilent MassHunter software version 10.1
Software version 10.1 p.2 ↗ Agilent MassHunter
Calibration range 0.01 to 0.5 μg/mL p.5 ↗ as shown in Figure 7
LLOQ ≤ 0.02 μg/mL p.5 ↗ obtained LLOQ
Number of analytes tested 100 p.2 ↗ SVOCs

Operating Limits

  • LLOQ: ≤ 0.02 μg/mL p.5 ↗
  • Calibration range: 0.01 to 0.5 μg/mL p.5 ↗