Ethylene Oxide and 2-Chloroethanol in Spices and Oilseeds Using QuEChERS GC/MS/MS
Technical Reference Summary
Covered Models & Document Scope
Application note for the determination of ethylene oxide (EtO) and 2-chloroethanol (2-CE) in spices and oilseeds using QuEChERS extraction with an Agilent 8890 GC / G4513 autosampler coupled to an Agilent 7010 triple quadrupole GC/MS system; covers GC/TQ method parameters and results for flaxseed, cumin powder and chili powder.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| GC system | Agilent 8890 GC with G4513 autosampler p.2 ↗ | |
| Mass spectrometer | Agilent 7010 triple quadrupole mass spectrometry system p.2 ↗ | |
| Analytical column | Agilent J&W DB-VRX (60 m x 0.25 mm, 1.4 µm) p.2 ↗ | |
| Column flow | Helium: 1.0 mL/min p.2 ↗ | constant flow |
| Injection mode | Pulsed split (4:1) p.2 ↗ | |
| Injection volume | 2 µL p.2 ↗ | |
| Injection program | Starts at 90 °C (hold for 0.8 min), ramped at 450 °C/min to 250 °C, hold for 10 min p.2 ↗ | |
| Oven program | Starts at 40 °C (hold for 1.0 min), ramped at 10 °C/min to 160 °C, and then at 30 °C/min to 245 °C, hold for 5 min p.2 ↗ | |
| Ionization mode | EI p.2 ↗ | |
| Ion source temperature | 230 °C p.2 ↗ | |
| Quadrupole temperature (Q1 and Q2) | 150 °C p.2 ↗ | |
| MRM transition EtO | 44 & 14 (CE:20) p.2 ↗ | |
| MRM transition 2-CE | 80 & 31 (CE:5) p.2 ↗ | |
| Stock standard concentration (EtO and 2-CE) | 1 mg/mL p.2 ↗ | each |
| Working standard concentration (EtO and 2-CE) | 10 µg/ml p.2 ↗ | |
| Calibration standards | 2.0, 5.0, 10.0, 20.0, 50.0 p.2 ↗ | |
| Standard preparation temperature | < 10 °C p.2 ↗ | |
| Centrifugation speed | 5,000 rpm p.2 ↗ | 5 minutes |
Operating Limits
- Standard preparation temperature: < 10 °C p.2 ↗
- Acetonitrile diluent freezer temperature: –20 °C p.2 ↗
- Minimum acetonitrile cooling time: 15 minutes p.2 ↗