Induztr
Engineering Digest Document Viewer Related Models

Ethylene Oxide and 2-Chloroethanol in Spices and Oilseeds Using QuEChERS GC/MS/MS

Technical Reference Summary

Covered Models & Document Scope

Application note for the determination of ethylene oxide (EtO) and 2-chloroethanol (2-CE) in spices and oilseeds using QuEChERS extraction with an Agilent 8890 GC / G4513 autosampler coupled to an Agilent 7010 triple quadrupole GC/MS system; covers GC/TQ method parameters and results for flaxseed, cumin powder and chili powder.

Technical Data

Parameter Value Condition
GC system Agilent 8890 GC with G4513 autosampler p.2 ↗
Mass spectrometer Agilent 7010 triple quadrupole mass spectrometry system p.2 ↗
Analytical column Agilent J&W DB-VRX (60 m x 0.25 mm, 1.4 µm) p.2 ↗
Column flow Helium: 1.0 mL/min p.2 ↗ constant flow
Injection mode Pulsed split (4:1) p.2 ↗
Injection volume 2 µL p.2 ↗
Injection program Starts at 90 °C (hold for 0.8 min), ramped at 450 °C/min to 250 °C, hold for 10 min p.2 ↗
Oven program Starts at 40 °C (hold for 1.0 min), ramped at 10 °C/min to 160 °C, and then at 30 °C/min to 245 °C, hold for 5 min p.2 ↗
Ionization mode EI p.2 ↗
Ion source temperature 230 °C p.2 ↗
Quadrupole temperature (Q1 and Q2) 150 °C p.2 ↗
MRM transition EtO 44 & 14 (CE:20) p.2 ↗
MRM transition 2-CE 80 & 31 (CE:5) p.2 ↗
Stock standard concentration (EtO and 2-CE) 1 mg/mL p.2 ↗ each
Working standard concentration (EtO and 2-CE) 10 µg/ml p.2 ↗
Calibration standards 2.0, 5.0, 10.0, 20.0, 50.0 p.2 ↗
Standard preparation temperature < 10 °C p.2 ↗
Centrifugation speed 5,000 rpm p.2 ↗ 5 minutes

Operating Limits

  • Standard preparation temperature: < 10 °C p.2 ↗
  • Acetonitrile diluent freezer temperature: –20 °C p.2 ↗
  • Minimum acetonitrile cooling time: 15 minutes p.2 ↗