Induztr
Engineering Digest Document Viewer Related Models

Fatty Acid Methyl Ester Analysis Using the Agilent 8850 GC System

Technical Reference Summary

Covered Models & Document Scope

Agilent application note on FAME analysis by GC using the Agilent 8850 GC system; covers Agilent J&W DB-FastFAME columns (20 m × 0.18 mm, 30 m × 0.25 mm) and Agilent J&W HP-88 column (100 m, 0.25 mm), with He and N2 carrier gas methods; no product ordering code or size matrix is given.

Technical Data

Parameter Value Condition
Column type Agilent J&W DB-FastFAME p.3 ↗ 20 m × 0.18 mm, He
Column head pressure 28 psi p.3 ↗ 20 m DB-FastFAME, CP
Oven temperature program 60 °C (0.49 min), 300 °C/min to 175 °C (0.49 min), 15 °C/min to 210 °C (2.5 min), 14.5 °C/min to 240 °C (3 min) °C p.3 ↗ 20 m DB-FastFAME, He
Inlet septa Hi-temp/low bleed/nonstick septa p.3 ↗
Inlet liner Ultra-inert, low pressure drop split liner p.3 ↗
Resolution 1.27 p.4 ↗ C18:1n9c / C18:2n6c, 20 m DB-FastFAME

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
Agilent J&W DB-FastFAME, 20 m × 0.18 mm p.3 ↗ 100-2000 Column 1
Agilent J&W DB-FastFAME, 30 m × 0.25 mm p.3 ↗ G3903-63011 Column 2
Agilent J&W HP-88 GC column, 100 m, 0.25 mm p.3 ↗ 112-88A7E Column 4
Hi-temp/low bleed/nonstick septa p.3 ↗ 5183-4757 Inlet septa
Ultra-inert, low pressure drop split liner p.3 ↗ 5190-2295 Inlet liner