Induztr
Engineering Digest Document Viewer Related Models

A Fast and Robust GC/MS/MS Analysis of 203 Pesticides in 10 Minutes in Spinach

Technical Reference Summary

Covered Models & Document Scope

Application note covering two fast GC/MS/MS methods for analysis of 203 pesticides in spinach: (a) 15 m × 15 m (0.25 mm × 0.25 µm) midcolumn backflush configuration and (b) minibore 10 m × 10 m (0.18 mm × 0.18 µm) midcolumn backflush configuration, on Agilent 8890 GC with 7000E or 7010C triple quadrupole GC/MS.

Technical Data

Parameter Value Condition
Analysis time 10 min p.1 ↗ 15 m × 15 m configuration
Analysis time 10 min p.1 ↗ 10 m × 10 m configuration
Number of pesticides analyzed 203 p.1 ↗
Calibration range 0.1 to 1,000 ppb p.1 ↗ 7010C and 7000 series
Calibration correlation coefficient R2 > 0.99 p.10 ↗ linear or quadratic fit
Number of consecutive injections (robustness) 700 injections p.1 ↗ spinach extract spiked at 20 ppb
Continuous running time (robustness) 175 hours p.1 ↗ 700 injections
Liner and septum replacement interval every 100 injections p.2 ↗
GC oven voltage 220 V p.4 ↗ Agilent 8890 GC
Column 1 length 15 m p.4 ↗ 15 m × 15 m configuration
Column 1 diameter 0.25 mm p.4 ↗ 15 m × 15 m configuration
Column 1 film thickness 0.25 µm p.4 ↗ 15 m × 15 m configuration
Injection volume 1.0 µL p.4 ↗
Inlet temperature 60 °C for 0.1 min, then 310 °C p.4 ↗
Initial oven temperature 60 °C p.4 ↗ 15 m × 15 m configuration
Initial oven temperature 60 °C p.4 ↗ 10 m × 10 m configuration
Ramp rate 1 80 °C/min p.4 ↗ 15 m × 15 m configuration
Ramp rate 1 80 °C/min p.4 ↗ 10 m × 10 m configuration
Final temperature 2 310 °C p.4 ↗ 15 m × 15 m configuration
Final temperature 2 310 °C p.4 ↗ 10 m × 10 m configuration
Total run time 10 min p.4 ↗ 15 m × 15 m configuration

Operating Limits

  • Minimum calibration concentration: 0.1 ppb p.1 ↗
  • Maximum calibration concentration: 1,000 ppb p.1 ↗
  • Maximum number of consecutive injections demonstrated: 700 injections p.1 ↗
  • Maximum continuous running time demonstrated: 175 hours p.1 ↗
  • Liner and septum replacement interval: every 100 injections p.2 ↗