Hydrogen Carrier Gas for Analyzing Pesticides in Pigmented Foods with GC/MS/MS
Technical Reference Summary
Covered Models & Document Scope
Application note covering GC/TQ pesticide analysis using hydrogen carrier gas with Agilent 8890 GC, 7000E and 7010C GC/TQ systems, HydroInert and HES EI sources, HP-5ms UI 20 m × 0.18 mm × 0.18 µm column, and 203 target pesticides in pigmented spinach matrix
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Column length | 20 m p.4 ↗ | Agilent HP-5ms UI (p/n 19091S-577UI) |
| Column diameter | 0.18 mm p.4 ↗ | Agilent HP-5ms UI |
| Film thickness | 0.18 µm p.4 ↗ | Agilent HP-5ms UI |
| Column flow | 1.0 mL/min p.4 ↗ | constant flow, nominal before retention time locking |
| Inlet temperature | 60 °C for 0.06 min, then to 280 °C at 60 °C p.4 ↗ | Multimode Inlet (MMI), solvent vent mode |
| Initial oven temperature | 60 °C p.4 ↗ | oven program |
| Oven ramp rate 2 | 10 °C/min p.4 ↗ | oven program |
| Total run time | 20 min p.4 ↗ | 10-minute analysis with backflush |
| Purge flow to split vent | 60 mL/min p.4 ↗ | at 2.56 min |
| Septum purge flow | 3 mL/min p.4 ↗ | switched mode |
| Vent flow | 100 mL/min p.4 ↗ | solvent vent mode |
| Vent pressure | 5 psi p.4 ↗ | until 0.06 min |
| Cryo use temperature | 200 °C p.4 ↗ | cryo on (air) |
| Total MRMs | 614 p.4 ↗ | dMRM mode |
| Minimum dwell time | 3 ms p.4 ↗ | dMRM mode |
| EM voltage gain mode | 10 p.4 ↗ | MSD |
| Vacuum pump | Performance turbo p.4 ↗ | Agilent 8890 GC/TQ system |
| Quantitation range | 0.1 to 1,000 ppb p.14 ↗ | 7010C GC/TQ with HES, quadratic fit |
| Default MRL | 10 ppb p.2 ↗ | spinach matrix |
| Target pesticides quantified at or below MRL | over 90% of 203 % p.2 ↗ | HydroInert and HES sources |
| RSE below 20% | over 94% of 203 % p.2 ↗ | SANTE/11312/2021 guidelines |
| Analyte response enhancement | 10-fold on average fold p.1 ↗ | optimized injection conditions vs helium injection conditions with hydrogen |
| MDL for reactive compounds | sub-ppb ppb p.2 ↗ | HES and HydroInert with hydrogen |
| Lowest detectable level | 0.5 ppb p.11 ↗ | HydroInert with H2 |
Operating Limits
- Maximum oven temperature: 310 °C p.8 ↗
- Minimum calibration concentration: 0.1 ppb p.14 ↗
- Maximum calibration concentration: 1,000 ppb p.14 ↗
- Default MRL: 10 ppb p.2 ↗