HILIC Analysis of GLP-1 Receptor Agonists
Technical Reference Summary
Covered Models & Document Scope
Application note comparing an Agilent 1290 Infinity III LC System (stainless-steel, SST) and a 1290 Infinity III Bio LC System (BIO) for HILIC analysis of GLP-1 receptor agonist drug products using Poroshell HILIC-Z and Altura Poroshell HILIC-Z columns with DAD and ELSD detection.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Column | Agilent InfinityLab Poroshell HILIC-Z, 2.1 × 150 mm, 2.7 µm (p/n 683775-924) p.3 ↗ | SST system |
| Flow rate | 0.4 mL/min p.3 ↗ | |
| Mobile phase A | 100 mM ammonium formate pH 3 p.3 ↗ | |
| Mobile phase B | ACN p.3 ↗ | |
| Column temperature | 40 °C p.3 ↗ | |
| Autosampler temperature | 10 °C p.3 ↗ | |
| Injection volume | 1 µL p.3 ↗ | |
| Needle wash | 10 s 50/50 (v/v) acetonitrile:water p.3 ↗ | |
| DAD detection wavelength | 280/4 nm, reference 360/100 nm, 2.5 Hz p.3 ↗ | |
| ELSD evaporator temperature | 50 °C p.3 ↗ | |
| ELSD nebulizer temperature | 50 °C p.3 ↗ | |
| ELSD gas flow rate | 1.0 SLM p.3 ↗ | |
| ELSD data rate | 80 Hz p.3 ↗ | |
| ELSD smoothing | 30 p.3 ↗ | |
| ELSD PMT gain | 1.0 p.3 ↗ | |
| Pump | Agilent 1290 Infinity III High-Speed Pump (G7120A) p.3 ↗ | SST system |
| Detection cell | Agilent Max-Light Cartridge Cell, standard p.3 ↗ | SST system |
| Data system | Agilent OpenLab CDS, version 2.7 p.3 ↗ |
Operating Limits
- Column temperature: 40 °C p.3 ↗
- Autosampler temperature: 10 °C p.3 ↗
- ELSD evaporator/nebulizer temperature: 50 °C p.3 ↗