Quantitative Analysis of 210 Veterinary Drugs in Organ Meat Using the Agilent 6470A Triple Quadrupole LC/MS
Technical Reference Summary
Covered Models & Document Scope
Application note covering the Agilent Comprehensive Veterinary Drug dMRM Solution (G5368AA) for screening 210 veterinary drug residues in organ meat (chicken kidney and liver) using the Agilent 6470 Triple Quadrupole LC/MS with 1290 Infinity II LC system; no instrument model variants or ordering codes are described.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Number of target residues | 210 p.1 ↗ | |
| Number of chemical classes covered | >28 p.1 ↗ | |
| Sample amount | 2.0 ±0.1 g p.2 ↗ | homogenized chicken kidney and liver |
| Retention time precision (%RSD) | <0.6 % p.3 ↗ | duplicate injections of technical preparations, each QC level |
| Response precision (%RSD) | <20 % p.3 ↗ | duplicate injections of technical preparations, each QC level |
| Accuracy | 70 to 120 % p.3 ↗ | kidney and liver QC levels |
| Target recovery | 60 to 120 % p.3 ↗ | |
| QC RSD | <20 % p.5 ↗ | |
| Calibration linearity R2 | >0.99 p.5 ↗ | LOD range |
| LOD range | 0.1–10 µg/kg p.5 ↗ | kidney and liver, per chemical class |
| MRL range | 0.1–2,000 µg/kg p.5 ↗ | kidney and liver, per chemical class |
| Total targets with established MRL | 91 p.5 ↗ | |
| Targets detected in kidney | 206 p.5 ↗ | |
| Targets detected in liver | 195 p.5 ↗ | |
| Targets meeting recovery criteria in kidney | 189 p.5 ↗ | |
| Targets meeting recovery criteria in liver | 206 p.5 ↗ | |
| Targets with negligible interference | about 92 % p.4 ↗ | MRL-based screening |
Operating Limits
- Retention time precision (%RSD): <0.6% p.3 ↗
- Response precision (%RSD): <20% p.3 ↗
- Accuracy: 70% to 120% p.3 ↗
- Target recovery: 60 to 120% p.3 ↗
- QC RSD: <20% p.5 ↗
- Calibration linearity R2: >0.99 p.5 ↗