Determination of Ultratrace Impurities in Semiconductor Photoresist Using ICP-MS/MS

Technical Reference Summary

Covered Models & Document Scope

Application note covering the Agilent 8900 #200 ICP-MS/MS (ICP-QQQ) configured with a 200 µL/min MicroFlow PFA nebulizer and O2/Ar option gas, used for determination of 20 elemental contaminants in IC photoresist after simple dilution in PGMEA solvent; tune modes NH3, H2 and O2 are covered.

Technical Data

Parameter Value Condition
Nebulizer flow rate 200 µL/min p.3 ↗ 8900 #200, MicroFlow PFA nebulizer
Option gas 100% O2 p.3 ↗ O2/Ar mix
RF power 1000 W p.4 ↗ NH3 tune mode
Sampling depth 18.0 mm p.4 ↗ NH3 tune mode
Makeup gas 0.60 L/min p.4 ↗ NH3 tune mode
O2/Ar option gas 0.25 L/min p.4 ↗ NH3 tune mode
Omega lens 5 V p.4 ↗ NH3 tune mode
Axial acceleration 0.5 V p.4 ↗ NH3 tune mode

Operating Limits

  • Maximum acceptable metallic impurity level in photoresist: 1–10 ppb p.6 ↗