Routine Trace‑Level Analysis of m‑Xylene in High‑Purity p‑Xylene using the Agilent 8890B GC System
Technical Reference Summary
Covered Models & Document Scope
Application note covering the Agilent 8890B GC system applied to trace-level m-xylene in high-purity p-xylene; two method variants: High-Resolution Method (50 m × 0.200 mm, 0.4 µm column, 19.3 min run) and High-Speed Method (20 m × 0.100 mm, 0.2 µm column, 13.83 min run).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Column dimensions | 50 m × 0.200 mm, 0.4 µm p.2 ↗ | High-Resolution Method |
| Detector | FID p.2 ↗ | High-Resolution Method |
| Detector | FID p.2 ↗ | High-Speed Method |
| Carrier gas | Hydrogen p.2 ↗ | High-Resolution Method |
| Carrier gas | Hydrogen p.2 ↗ | High-Speed Method |
| Run time | 19.3 minutes p.3 ↗ | High-Resolution Method |
| Septum purge | 6 mL/min p.3 ↗ | High-Resolution Method |
| Septum purge | 6 mL/min p.3 ↗ | High-Speed Method |
| Injection volume | 0.3 µL p.3 ↗ | High-Resolution Method |
| Injection volume | 0.3 µL p.3 ↗ | High-Speed Method |
| Inlet mode | Split, 250:1 p.3 ↗ | High-Resolution Method |
| Inlet temperature | 250 °C p.3 ↗ | High-Resolution Method |
| Oven initial temperature | 40 °C p.3 ↗ | High-Resolution Method |
| Oven initial temperature | 40 °C p.3 ↗ | High-Speed Method |
| Oven ramp 1 rate | 3 °C/min p.3 ↗ | High-Resolution Method |
| Oven ramp 1 rate | 3 °C/min p.3 ↗ | High-Speed Method |
| Oven ramp 2 rate | 50 °C/min p.3 ↗ | High-Resolution Method |
| Oven ramp 2 rate | 50 °C/min p.3 ↗ | High-Speed Method |
| Oven ramp 2 hold | 1 minute p.3 ↗ | High-Resolution Method |
Operating Limits
- Linearity range for m-xylene: 10–50 ppm p.1 ↗
- R², High-Resolution Method: 0.99950 p.1 ↗
- R², High-Speed Method: 0.99975 p.1 ↗
- Number of consecutive injections for precision evaluation: 200 p.1 ↗
- m-xylene concentration for precision evaluation: 50 ppm p.1 ↗