Routine Trace‑Level Analysis of m‑Xylene in High‑Purity p‑Xylene using the Agilent 8890B GC System

Technical Reference Summary

Covered Models & Document Scope

Application note covering the Agilent 8890B GC system applied to trace-level m-xylene in high-purity p-xylene; two method variants: High-Resolution Method (50 m × 0.200 mm, 0.4 µm column, 19.3 min run) and High-Speed Method (20 m × 0.100 mm, 0.2 µm column, 13.83 min run).

Technical Data

Parameter Value Condition
Column dimensions 50 m × 0.200 mm, 0.4 µm p.2 ↗ High-Resolution Method
Detector FID p.2 ↗ High-Resolution Method
Detector FID p.2 ↗ High-Speed Method
Carrier gas Hydrogen p.2 ↗ High-Resolution Method
Carrier gas Hydrogen p.2 ↗ High-Speed Method
Run time 19.3 minutes p.3 ↗ High-Resolution Method
Septum purge 6 mL/min p.3 ↗ High-Resolution Method
Septum purge 6 mL/min p.3 ↗ High-Speed Method
Injection volume 0.3 µL p.3 ↗ High-Resolution Method
Injection volume 0.3 µL p.3 ↗ High-Speed Method
Inlet mode Split, 250:1 p.3 ↗ High-Resolution Method
Inlet temperature 250 °C p.3 ↗ High-Resolution Method
Oven initial temperature 40 °C p.3 ↗ High-Resolution Method
Oven initial temperature 40 °C p.3 ↗ High-Speed Method
Oven ramp 1 rate 3 °C/min p.3 ↗ High-Resolution Method
Oven ramp 1 rate 3 °C/min p.3 ↗ High-Speed Method
Oven ramp 2 rate 50 °C/min p.3 ↗ High-Resolution Method
Oven ramp 2 rate 50 °C/min p.3 ↗ High-Speed Method
Oven ramp 2 hold 1 minute p.3 ↗ High-Resolution Method

Operating Limits

  • Linearity range for m-xylene: 10–50 ppm p.1 ↗
  • R², High-Resolution Method: 0.99950 p.1 ↗
  • R², High-Speed Method: 0.99975 p.1 ↗
  • Number of consecutive injections for precision evaluation: 200 p.1 ↗
  • m-xylene concentration for precision evaluation: 50 ppm p.1 ↗