Induztr
Engineering Digest Document Viewer Related Models

Analysis of Trace Impurities in Monocyclic Aromatic Solvents Following ASTM D7504

Technical Reference Summary

Covered Models & Document Scope

Application note covering a dual-channel Agilent 8860B gas chromatograph (GC/FID) system for ASTM D7504 analysis of trace impurities in monocyclic aromatic solvents (benzene, toluene, xylenes, styrene). Three carrier-gas methods are covered: helium, nitrogen, and hydrogen.

Technical Data

Parameter Value Condition
Injection volume 0.6 µL p.3 ↗
Inlet mode Split, 100:1 p.3 ↗
Inlet temperature 260 °C p.3 ↗
Oven initial temperature 60 °C p.3 ↗ He, H2 and N2 methods
Oven initial hold 10 min p.3 ↗ He and H2 methods
Ramp 1 rate 5 °C/min p.3 ↗ He and H2 methods
Ramp 1 set point 150 °C p.3 ↗ He, H2 and N2 methods
Ramp 2 rate 30 °C/min p.3 ↗ He, H2 and N2 methods
Ramp 2 set point 220 °C p.3 ↗ He, H2 and N2 methods
Ramp 2 hold 8 min p.3 ↗ He, H2 and N2 methods
Post run temperature 0 °C p.3 ↗ He, H2 and N2 methods
Average signal-to-noise ratio 43:1 p.10 ↗ He method, front channel
Average area 0.838 p.10 ↗ He method, front channel

Operating Limits

  • Mass concentration precision (RSD) for most impurities: below 1% p.1 ↗
  • Reproducibility limit (R) for benzene: 0.0247 p.9 ↗
  • Reproducibility limit (R) for toluene: 0.011 p.9 ↗
  • Reproducibility limit (R) for ethylbenzene: 0.001 p.9 ↗
  • Reproducibility limit (R) for p-xylene: 0.0173 p.9 ↗
  • Reproducibility limit (R) for m-xylene: 0.0165 p.9 ↗