Analysis of Trace Impurities in Monocyclic Aromatic Solvents Following ASTM D7504
Technical Reference Summary
Covered Models & Document Scope
Application note covering a dual-channel Agilent 8860B gas chromatograph (GC/FID) system for ASTM D7504 analysis of trace impurities in monocyclic aromatic solvents (benzene, toluene, xylenes, styrene). Three carrier-gas methods are covered: helium, nitrogen, and hydrogen.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Injection volume | 0.6 µL p.3 ↗ | |
| Inlet mode | Split, 100:1 p.3 ↗ | |
| Inlet temperature | 260 °C p.3 ↗ | |
| Oven initial temperature | 60 °C p.3 ↗ | He, H2 and N2 methods |
| Oven initial hold | 10 min p.3 ↗ | He and H2 methods |
| Ramp 1 rate | 5 °C/min p.3 ↗ | He and H2 methods |
| Ramp 1 set point | 150 °C p.3 ↗ | He, H2 and N2 methods |
| Ramp 2 rate | 30 °C/min p.3 ↗ | He, H2 and N2 methods |
| Ramp 2 set point | 220 °C p.3 ↗ | He, H2 and N2 methods |
| Ramp 2 hold | 8 min p.3 ↗ | He, H2 and N2 methods |
| Post run temperature | 0 °C p.3 ↗ | He, H2 and N2 methods |
| Average signal-to-noise ratio | 43:1 p.10 ↗ | He method, front channel |
| Average area | 0.838 p.10 ↗ | He method, front channel |
Operating Limits
- Mass concentration precision (RSD) for most impurities: below 1% p.1 ↗
- Reproducibility limit (R) for benzene: 0.0247 p.9 ↗
- Reproducibility limit (R) for toluene: 0.011 p.9 ↗
- Reproducibility limit (R) for ethylbenzene: 0.001 p.9 ↗
- Reproducibility limit (R) for p-xylene: 0.0173 p.9 ↗
- Reproducibility limit (R) for m-xylene: 0.0165 p.9 ↗