Analyse ausgewählter Alkohole, Aldehyde und Ester in destillierten
Technical Reference Summary
Covered Models & Document Scope
Application Note: analysis of selected aroma compounds (alcohols, aldehydes, esters) in distilled spirits (Chinese schnaps) using an Agilent 8860 GC system with FID detector and Agilent 7693A automatic liquid sampler (ALS) with 5-µl syringe; quantitative analysis of the 10 most common compounds.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| GC system | 8860 p.2 ↗ | |
| Detector | FID p.2 ↗ | |
| Autosampler | 7693A p.2 ↗ | |
| Syringe volume | 5 µl p.2 ↗ | |
| Standard stock solution concentration | 1000 µg/ml p.2 ↗ | 10 compounds, 60:40 ethanol:water (v/v) |
| Oven temperature ramp | 5 °C/min p.2 ↗ | up to 100 °C |
| Calibration concentration | 100 µg/ml p.5 ↗ | |
| Retention time, Ethylacetat | 4,51 min p.5 ↗ | |
| Retention time, Methanol | 4,75 min p.5 ↗ | |
| Retention time, 2-Butanol | 7,88 min p.5 ↗ | |
| Retention time, Ethylbutyrat | 8,09 min p.5 ↗ | |
| Retention time, n-Propanol | 8,26 min p.5 ↗ | |
| Retention time, Isobutanol | 9,96 min p.5 ↗ | |
| Retention time, 1-Butanol | 11,43 min p.5 ↗ | |
| Retention time, Isoamylalkohol | 13,25 min p.5 ↗ | |
| Retention time, Ethylhexanoat | 13,88 min p.5 ↗ | |
| Retention time, Ethyllactat | 16,93 min p.5 ↗ |