Verwendung eines Pneumatikschaltmoduls (PSD)

Technical Reference Summary

Covered Models & Document Scope

Application note (German) describing a backflush configuration on an Agilent 8890 GC system with MMI inlet, PSD module and FPD Plus detector; covers the 8890 GC and mentions the Intuvo 9000 GC as an alternative system. No product ordering codes or size variants are defined.

Technical Data

Parameter Value Condition
Gas chromatograph 8890 series GC p.3 ↗
Inlet type MMI p.3 ↗
MMI program 100 °C (0.02 min), 900 °C/min to 450 °C p.3 ↗
Oven (after run) 360 °C (5 minutes) p.3 ↗
PSD pressure 70 psi p.3 ↗ 4.5 ml/min column 2/3
Inlet pressure 2 psi p.3 ↗ 4.5 ml/min column 1
Column Agilent J&W DB-HT Sim Dis, 5 m × 530 µm p.3 ↗
Carrier gas flow 5 ml/min p.3 ↗ helium, constant flow mode
Liner Ultra Inert, split, glass wool p.3 ↗
Syringe Blue Line, 5 µl, conical p.3 ↗
Column 1 J&W DB-1ms UI p.3 ↗
Purge flow setpoint range 3 to 30 ml/min p.2 ↗ PSD purge flow
Purge flow default setpoint 3 ml/min p.2 ↗ PSD purge flow

Operating Limits

  • Purge flow adjustable range: 3 to 30 ml/min p.2 ↗
  • Purge flow default setpoint: 3 ml/min p.2 ↗