Induztr
Engineering Digest Document Viewer Related Models

Utilisation du PSD pour le rétrobalayage du GC Agilent 8890

Technical Reference Summary

Covered Models & Document Scope

Agilent 8890 GC with PSD (pneumatic switching device) EPC module and Deans switch backflush configuration; application note, no product variants or ordering codes.

Technical Data

Parameter Value Condition
Purge flow range 3 to 30 mL/min p.2 ↗ PSD purge flow setting
Purge flow default 3 mL/min p.2 ↗ PSD purge flow setting
PSD pressure 70 psi p.3 ↗ 4.5 mL/min column 2/3
Inlet type MMI p.3 ↗
MMI program 100 °C (0.02 minute), 900 °C/min jusqu'à p.3 ↗
Column DB-HT Sim Dis Agilent J&W, 5 m × 530 µm p.3 ↗
Second column DB-17ht Agilent J&W (réf. 122-1831) p.3 ↗
Syringe Gamme bleue, 5 µL, conique p.3 ↗
Inlet insert Ultra-inerte, division, laine de verre p.3 ↗
Carrier gas hélium p.2 ↗ constant flow mode
Deans switch isolation window 20.1 to 20.4 min p.2 ↗

Operating Limits

  • PSD purge flow setting range: 3 to 30 mL/min p.2 ↗
  • PSD purge flow default: 3 mL/min p.2 ↗