Bestimmung vom Benzol und seinen Derivaten in Wasser mit dem

Technical Reference Summary

Covered Models & Document Scope

Application note for the determination of benzene and its derivatives in water by headspace GC-FID according to method HJ 1067-2019, using an Agilent 8697 Headspace Sampler coupled to an Agilent 8890 GC/FID system with Agilent OpenLab CDS 2.5.

Technical Data

Parameter Value Condition
Number of target compounds 8 p.2 ↗ benzene, toluene, ethylbenzene, p-xylene, m-xylene, cumene, o-xylene, styrene
Calibration concentration levels 10, 20, 50, 200, 500, 2000 µg/l p.2 ↗
Stock solution concentration 1000 µg/ml p.2 ↗ in methanol
Intermediate stock solution concentrations 10 and 100 µg/ml p.2 ↗ in methanol
Sodium chloride per vial 3 g p.2 ↗
Reagent water per vial 10 ml p.2 ↗ ultrapure water
Correlation coefficient R² > 0.999 p.2 ↗ all compounds
RSD of peak area 1.3 to 2.4 % p.2 ↗
RSD of retention time < 0.045 % p.2 ↗
Method detection limit (MDL) ≤ 0.2 µg/l p.2 ↗ all compounds
Recovery 99.1 to 101.7 % p.2 ↗
Retention time, benzene 4.839 min p.4 ↗
Retention time, toluene 7.807 min p.4 ↗
Retention time, ethylbenzene 10.771 min p.4 ↗
Retention time, p-xylene 10.994 min p.4 ↗
Retention time, m-xylene 12.068 min p.4 ↗
Retention time, cumene 12.463 min p.4 ↗
Retention time, o-xylene 15.173 min p.4 ↗
MDL, benzene 0.16 µg/l p.4 ↗
MDL, toluene 0.14 µg/l p.4 ↗
MDL, ethylbenzene 0.16 µg/l p.4 ↗
MDL, p-xylene 0.16 µg/l p.4 ↗
MDL, m-xylene 0.17 µg/l p.4 ↗
MDL, cumene 0.16 µg/l p.4 ↗
MDL, o-xylene 0.16 µg/l p.4 ↗

Operating Limits

  • Method detection limit (MDL) for all compounds: ≤ 0.2 µg/l p.2 ↗
  • RSD of peak area: 1.3 to 2.4 % p.2 ↗
  • RSD of retention time: < 0.045 % p.2 ↗
  • Recovery range: 99.1 to 101.7 % p.2 ↗