Biogas-Analyzer auf Basis des Agilent 990 Micro GC
Technical Reference Summary
Covered Models & Document Scope
Agilent 990 Micro GC Biogas-Analyzer, based on the 990 Micro GC platform (derived from 490 models). Two analyzer types: 990 Micro GC Biogas-Analyzer (two analytical channels: 10 m Agilent J&W CP-Molesieve 5Å with backflush, and 10 m Agilent J&W CP-PoraPLOT U with backflush) and an extended Agilent 990 Micro GC Biogas-Analyzer with a third direct channel (6 m Agilent J&W CP-Sil 5 CB). Base housing holds up to two analytical channels; two housings can be combined for up to four channels.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Analytical channels | bis zu zwei p.2 ↗ | 990 Micro GC base housing |
| Column 1 | 10 m Agilent J&W CP-Molesieve 5Å, Backflush, RTS p.2 ↗ | Biogas-Analyzer, channel 1 |
| Column 2 | 10 m Agilent J&W CP-PoraPLOT U, Backflush p.2 ↗ | Biogas-Analyzer, channel 2 |
| Column 3 | 6 m Agilent J&W CP-Sil 5 CB, direct p.2 ↗ | Extended Biogas-Analyzer, channel 3 |
| Carrier gas | Argon p.2 ↗ | channel 1 |
| Column head pressure | 200 kPa p.2 ↗ | channel 1 |
| Column temperature | 80 °C p.2 ↗ | channel 1 |
| Injection time | 40 ms p.2 ↗ | channel 1 |
| Injection time | 40 ms p.2 ↗ | channel 2 |
| Injection time | 40 ms p.2 ↗ | channel 3 |
| Backflush time | 10 s p.2 ↗ | channel 1 |
| Signal inversion | Ja p.2 ↗ | channel 1 |
| Conditioning temperature | 110 °C p.2 ↗ | channel 1 |
| Conditioning temperature | 110 °C p.2 ↗ | channel 2 |
Operating Limits
- max analytical channels in base housing: 2 p.2 ↗
- max analytical channels with two housings combined: 4 p.2 ↗