Blood Alcohol Analysis with the Integrated Agilent 8697 Headspace Sampler on 8890 GC-Dual FID System

Technical Reference Summary

Covered Models & Document Scope

Agilent application note 5994-3126EN describing blood alcohol concentration (BAC) analysis by headspace GC using an Agilent 8697 headspace sampler with an Agilent 8890 GC (split/splitless inlet, dual FID) and J&W DB-BAC1 Ultra Inert / DB-BAC2 Ultra Inert columns; includes consumables and standards part numbers and resolution/linearity/reproducibility results.

Technical Data

Parameter Value Condition
GC system 8890 p.2 ↗ dual FID, split/splitless inlet
Headspace sampler 8697 p.1 ↗
Detector dual flame ionization detector (FID) p.2 ↗
Sample volume in headspace vial 0.5 mL p.2 ↗
Internal standard concentration 0.1 % (v/v) p.2 ↗ t-butanol
GC liner inner diameter 2 mm p.2 ↗ p/n 5190-6168
Vial volume 20 mL p.2 ↗ crimp caps
Number of sequential injections 12 p.4 ↗
Resolution, methanol 2.7 p.4 ↗ BAC1 UI
Resolution, acetaldehyde 7.9 p.4 ↗ BAC1 UI
Resolution, ethanol 2.0 p.4 ↗ BAC1 UI
Resolution, isopropanol 7.2 p.4 ↗ BAC1 UI
Resolution, t-butanol 5.2 p.4 ↗ BAC1 UI
Resolution, propanal 1.3* p.4 ↗ BAC1 UI, manually calculated
Resolution, n-propanol 2.0* p.4 ↗ BAC1 UI, manually calculated
Resolution, acetone 3.8 p.4 ↗ BAC1 UI
Resolution, acetonitrile 3.9 p.4 ↗ BAC1 UI
Resolution, 2-butanol 3.0 p.4 ↗ BAC1 UI
Resolution, ethyl acetate 12.4 p.4 ↗ BAC1 UI
Resolution, 2-butanone 4.1 p.4 ↗ BAC1 UI
Retention time, ethanol 1.348 min p.4 ↗ BAC1 UI
Retention time, t-butanol 1.953 min p.4 ↗ BAC1 UI

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
5190-2286 p.2 ↗ 5190-2286 20 mL vials and crimp caps
5190-6168 p.2 ↗ 5190-6168 GC liner (2 mm id)
5190-9765 p.2 ↗ 5190-9765 BAC resolution mix
G3440-24127 p.2 ↗ G3440-24127 Ethanol calibration standard