Application note covering analysis of BSA multimers using the Agilent 6560 ion mobility LC/Q-TOF with an Agilent nanospray source (G1992A); solution conditions: 150 mM ammonium acetate, 0.1% formic acid, and 150 mM ammonium acetate (per text). No product model/size variants or ordering codes are given.

Technical Data

Parameter Value Condition
Instrument 6560 ion mobility LC/Q-TOF p.1 ↗
Ionization source Agilent nanospray source (G1992A) p.2 ↗
Sample BSA (Sigma-Aldrich) p.2 ↗
Sample concentration 20 mg/mL p.2 ↗
Flow rate 266 nL/min p.2 ↗
Needle dimensions 360 µm od, 75 µm id, 30 µm tip, 5 cm length p.2 ↗ New Objective noncoated needle
Mobile phase 1 150 mM ammonium acetate p.2 ↗
Mobile phase 2 0.1% formic acid p.2 ↗
Drift tube length ~80 cm p.2 ↗
High pressure funnel voltage 200 V peak-to-peak p.3 ↗ optimized for native BSA transmission
Trapping funnel voltage 200 V peak-to-peak p.3 ↗ optimized for native BSA transmission
Monomer m/z (charge 1) 66,435 m/z p.4 ↗
Dimer m/z (charge 1) 132,869 m/z p.4 ↗
Trimer m/z (charge 1) 199,303 m/z p.4 ↗
Tetramer m/z (charge 1) 265,737 m/z p.4 ↗