Determination of Cd, Cr, Ni and Pb in Grains using the Agilent 280Z Zeeman GFAAS

Technical Reference Summary

Covered Models & Document Scope

Application brief for the determination of Cd, Cr, Ni and Pb in grains (wheat flour) using the Agilent 280Z Zeeman Graphite Furnace AAS with GTA 120 Graphite Tube Atomizer, UltrAA lamps and PSD 120 autosampler; covers method performance data only, no instrument model/size variants or ordering codes.

Technical Data

Parameter Value Condition
Method Detection Limit (MDL), Pb 2.6 ppb p.2 ↗ n=7
Method Detection Limit (MDL), Cd 0.37 ppb p.2 ↗ n=7
Method Detection Limit (MDL), Cr 1.5 ppb p.2 ↗ n=7
Method Detection Limit (MDL), Ni 6.0 ppb p.2 ↗ n=7
Certified value, Pb (NIST 1567b Wheat Flour) 10.4 ± 2.4 ppb p.2 ↗
Measured value, Pb 10.9 ppb p.2 ↗ mean, n=3
Standard deviation, Pb 1.4 ppb p.2 ↗ n=3
Recovery, Pb 105 % p.2 ↗
Certified value, Cd (NIST 1567b Wheat Flour) 25.4 ± 0.9 ppb p.2 ↗
Measured value, Cd 23.4 ppb p.2 ↗ mean, n=3
Standard deviation, Cd 0.22 ppb p.2 ↗ n=3
Recovery, Cd 92 % p.2 ↗
Certified value, Cr (DUWF-1 Durum Wheat Flour) 23.0 ± 9.0 ppb p.2 ↗
Measured value, Cr 21.5 ppb p.2 ↗ mean, n=3
Standard deviation, Cr 2.1 ppb p.2 ↗ n=3
Recovery, Cr 93 % p.2 ↗
Certified value, Ni (DUWF-1 Durum Wheat Flour) 170 ± 80 ppb p.2 ↗
Measured value, Ni 172 ppb p.2 ↗ mean, n=3
Standard deviation, Ni 7.5 ppb p.2 ↗ n=3
Recovery, Ni 101 % p.2 ↗