Analyse von Trinkwasser mit dem Agilent 8860 GC und dem 7697A

Technical Reference Summary

Covered Models & Document Scope

Agilent application note 5994-1239DEE (German, 15 October 2019): analysis of volatile halogenated and aromatic hydrocarbons and organophosphorus pesticides in drinking water using an Agilent 8860 GC with 7697A headspace sampler, FPD and Agilent DB-624 UI column (Best.-Nr. 122-1334UI); no instrument model/size variants or ordering codes are covered.

Technical Data

Parameter Value Condition
Headspace oven temperature 100 °C p.3 ↗
Carrier gas Stickstoff p.3 ↗
Vial purge flow 60 ml/min p.3 ↗
Split flow 50 ml/min p.3 ↗
Fuel gas air flow 60 ml/min p.3 ↗ FPD
Fuel gas H2 flow 60 ml/min p.3 ↗ FPD
Column Agilent DB-624 UI (Best.-Nr. 122-1334UI) p.3 ↗ Säule 1
Standard concentration 100 μg/ml p.4 ↗
Calibration levels 200, 500 und 1000 µg/l p.4 ↗
Area reproducibility (RSD, n = 10) 1,54 % p.5 ↗ Chloroform
Area reproducibility (RSD, n = 6) 1,36 % p.5 ↗ Dichlorvos

Operating Limits

  • Minimum calibration coefficient R2 for all compounds: > 0,9995 p.5 ↗
  • Maximum area RSD (n = 6): < 1,83 % p.5 ↗
  • Maximum retention time RSD (n = 10): < 0,017 % p.5 ↗