Analyse von Fettsäuremethylestern (FAMEs) auf einem Agilent 8890
Technical Reference Summary
Covered Models & Document Scope
Application note for FAME analysis on an Agilent 8890A GC system with split/splitless inlet and FID, using an Agilent 7693A ALS (G4567A); covers GB 5009.168-2016 (37-component FAME mix) and GB 5009.257-2016 (trans-FAME, 13-component mix and 8-component cis/trans C18:3 methyl ester mix) methods and real oil samples (soybean, peanut, sesame).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| GC system | 8890A GC p.2 ↗ | |
| Inlet | S/SL p.2 ↗ | |
| Inlet temperature | 250 °C p.2 ↗ | |
| Split ratio | 100:1 p.2 ↗ | |
| Liner | Split, Ultra inert, glass wool, low pressure p.2 ↗ | |
| Autosampler | 7693A ALS p.2 ↗ | Order No. G4567A |
| Detector | FID p.2 ↗ | |
| Carrier gas mode | constant pressure p.3 ↗ | |
| Oven temperature program | as recommended in GB 5009.168-2016 p.3 ↗ | |
| Analysis time | over 80 min p.3 ↗ | including injection |
| Sample solvent | hexane p.3 ↗ | |
| Sample concentration | 50–100 ng/µl p.3 ↗ | diluted |
| FAME mix concentration range | 200–400 ng/μl p.2 ↗ | 37-component mix, C4–C24 |
| Component weight percent range | 3–30 % p.2 ↗ | trans-FAME and cis/trans C18:3 mixes |
| RF RSD | only 4 % p.5 ↗ | C18:1-cis and C18:2-cis, five concentration levels |
| Maximum component deviation | 4.0 % p.3 ↗ |
Operating Limits
- Inlet temperature: 250 °C p.2 ↗
- Split ratio: 100:1 p.2 ↗
- Analysis time: over 80 min p.3 ↗
- Sample concentration: 50–100 ng/µl p.3 ↗
Ordering Data
| Ordering Code | Part No. | Variant / Code | Applies To | Note |
|---|---|---|---|---|
| Agilent 7693A automatic liquid sampler (ALS) p.2 ↗ | G4567A | Order number printed in materials list |