Forensische Analyse von Drogen mit dem Agilent 8890 GC-System
Technical Reference Summary
Covered Models & Document Scope
Application note (German) on forensic drug analysis using an Agilent 8890 GC system with split/splitless inlet, 5977 GC/MSD and 7693A ALS; covers method parameters and a 28-compound standard (Agilent GC/MS forensic toxicology mix, order no. 5190-0471). No product model/size variants or ordering codes are defined.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Syringe size | 10 µl p.2 ↗ | 8890 GC / 7693A ALS |
| Injection volume | 1 µl p.2 ↗ | 8890 GC |
| Inlet type | Split/Splitless p.2 ↗ | 8890 GC |
| Inlet mode | Split p.2 ↗ | 8890 GC |
| Inlet temperature | 250 °C p.2 ↗ | 8890 GC |
| Split flow | 30 ml/min p.2 ↗ | 8890 GC |
| Split ratio | 20:1 p.2 ↗ | 8890 GC |
| Gas saver time | Aus p.2 ↗ | 8890 GC |
| MS source | Extractor, 9-mm lens p.2 ↗ | 5977 GC/MSD |
| High vacuum pump | Turbo p.2 ↗ | 5977 GC/MSD |
| Acquisition mode | Scan p.2 ↗ | 5977 GC/MSD |
| Scan range | m/z 40 bis 500 m/z p.2 ↗ | 5977 GC/MSD |
| Tuning algorithm | Etune p.2 ↗ | 5977 GC/MSD |
| Source temperature | 250 °C p.2 ↗ | 5977 GC/MSD |
| Quadrupole temperature | 175 °C p.2 ↗ | 5977 GC/MSD |
| Standard concentration | 5 µg/ml p.2 ↗ | Agilent GC/MS forensic toxicology test mix, 28 compounds |
| Number of compounds in standard | 28 p.2 ↗ | Agilent GC/MS forensic toxicology test mix |
| GC/MS flow requirement (EPC capability) | unter 2 ml/min p.2 ↗ | 8890 GC, 6th-generation EPC |
Operating Limits
- GC/MS flow requiring pressure control (EPC capability): unter 2 ml/min p.2 ↗
Ordering Data
| Ordering Code | Part No. | Variant / Code | Applies To | Note |
|---|---|---|---|---|
| Agilent GC/MS Testmischung für forensische Toxikologie p.2 ↗ | 5190-0471 | 28 compounds, 5 µg/ml | Standard used in the study | |
| Advanced Green Septa p.2 ↗ | 5183-4761 | Consumable listed in the workflow table | ||
| Extraktorlinse, 9 mm p.2 ↗ | G3870-20449 | Consumable listed in the workflow table |