A New AdvanceBio HIC Column for Drug to Antibody Ratio (DAR) Analysis of Antibody Drug Conjugates (ADC)
Technical Reference Summary
Covered Models & Document Scope
Application note for DAR analysis of cysteine-linked antibody-drug conjugates (ADCs) by HIC using the Agilent 1260 Infinity II bio-inert LC system (pump G5654A, multisampler G5668A with sample cooler option 100, multicolumn thermostat G7116A with bio-inert heat exchanger option 019, diode array detector WR G7115A with bio-inert flow cell option 028) with AdvanceBio HIC columns 4.6 × 100 mm (p/n 685975-908) and 4.6 × 30 mm (p/n 681975-908); covers DAR 0, DAR 2, DAR 4, DAR 6 and DAR 8 variants.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Column | AdvanceBio HIC 4.6 × 100 mm mm p.2 ↗ | p/n 685975-908 |
| Mobile phase A | 50 mM Sodium phosphate, pH 7.0 p.2 ↗ | |
| Mobile phase B | 2 M Ammonium sulfate in 50 mM sodium phosphate, pH 7.0 p.2 ↗ | |
| Mobile phase C | Propan-2-ol p.2 ↗ | |
| Mobile phase D | HPLC grade water p.2 ↗ | |
| Flow rate | 0.5 mL/min p.2 ↗ | |
| Column temperature | 25 °C p.2 ↗ | |
| Injection volume | 5 µL p.2 ↗ | |
| Total run time | 31 minutes p.2 ↗ | |
| Gradient time 0 | 50 / 45 / 5 %A / %B / %C p.2 ↗ | typical gradient profile |
| Gradient time 15 | 75 / 0 / 25 %A / %B / %C p.2 ↗ | typical gradient profile |
| Gradient time 20 | 75 / 0 / 25 %A / %B / %C p.2 ↗ | typical gradient profile |
| Gradient time 21 | 50 / 45 / 5 %A / %B / %C p.2 ↗ | typical gradient profile |
| Gradient time 31 | 50 / 45 / 5 %A / %B / %C p.2 ↗ | typical gradient profile |
| Software | Agilent OpenLab 2.2 CDS p.2 ↗ |
Operating Limits
- column temperature: 25 °C p.2 ↗
- flow rate: 0.5 mL/min p.2 ↗
- injection volume: 5 µL p.2 ↗
- total run time: 31 minutes p.2 ↗
Ordering Data
| Ordering Code | Part No. | Variant / Code | Applies To | Note |
|---|---|---|---|---|
| AdvanceBio HIC 4.6 × 100 mm p.2 ↗ | 685975-908 | HIC column | ||
| AdvanceBio HIC 4.6 × 30 mm p.2 ↗ | 681975-908 | HIC column |