Determination of Epoxy Resin Polymers by GPC with the Agilent 1290 Infinity II LC
Technical Reference Summary
Covered Models & Document Scope
Application note describing micro-GPC determination of epoxy resin molecular weight using an Agilent 1290 Infinity II LC with 1290 Infinity II Micro-RID (G7162A) or 1260 Infinity II RID (G7162B); covers two column configurations (1 column MesoPore 2.1 × 250 mm PL1913-5325; 2 columns PLgel 3 µm MiniMixE 4.6 × 250 mm PL1510-5300) at flow rates 0.06–1.0 mL/min
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Solvent | THF, isocratic, Channel B p.2 ↗ | |
| Flow rate | 0.06 mL/min p.2 ↗ | 1 column, stop time 23 min |
| Column temperature | 35 °C p.2 ↗ | two columns in series connected with 75 µm capillary, 105 mm length |
| Injection volume | 20 µL p.2 ↗ | on 4.6 mm column |
| RID optical unit temperature | 35 °C p.2 ↗ | |
| Data rate | 9.25 Hz p.2 ↗ | |
| Signal polarity | Positive p.2 ↗ |
Operating Limits
- Column temperature: 35 °C p.2 ↗
- RID optical unit temperature: 35 °C p.2 ↗