High Sensitivity Determination of Lead in Soy Sauce

Technical Reference Summary

Covered Models & Document Scope

Application note for the high-sensitivity determination of lead in soy sauce using the Agilent 280Z Graphite Furnace Atomic Absorption Spectrometer (GFAAS) with Omega platform tube and PSD 120 autosampler; no instrument model/size variants or ordering codes are covered.

Technical Data

Parameter Value Condition
Instrument 280Z AA p.1 ↗ GFAAS
Background correction Transverse Zeeman p.2 ↗
Autosampler PSD 120 p.2 ↗
Graphite tube Omega platform tube p.2 ↗
Furnace viewing camera Tube-CAM p.2 ↗ standard on 280Z AA
Software Spectra 5.5 p.2 ↗ SRM Wizard
Drying step 1 temperature 95 °C p.3 ↗ 50.0 s, Ar 0.3 L/min
Drying step 2 temperature 120 °C p.3 ↗ 20.0 s, Ar 0.3 L/min
Ashing step 3 temperature 400 °C p.3 ↗ 30.0 s, Ar 0.3 L/min
Ashing step 4 temperature 500 °C p.3 ↗ 20.0 s, Ar 0.3 L/min
Ashing step 5 temperature 805 °C p.3 ↗ 15.0 s, Ar 0.3 L/min
Ashing step 6 temperature 805 °C p.3 ↗ 10.0 s, Ar 0.3 L/min
Ashing step 7 temperature 805 °C p.3 ↗ 1.0 s, Ar 0.0 L/min
Atomization step 8 temperature 2132 °C p.3 ↗ 0.6 s, Ar 0.0 L/min, reading
Atomization step 9 temperature 2132 °C p.3 ↗ 2.0 s, Ar 0.0 L/min, reading
Cleanout step 10 temperature 2500 °C p.3 ↗ 2.0 s, Ar 0.3 L/min
Ash temperature (SRM) 1000 °C p.3 ↗ step 300 °C
Atomization temperature (SRM) 2000 °C p.3 ↗ step 300 °C
MDL in solution 0.35 μg/kg p.4 ↗
MDL in sample 18 μg/kg p.4 ↗
LOQ in sample 59 μg/kg p.4 ↗
Maximum allowable limit (Pb in condiments) 1000 μg/kg p.4 ↗ GB 2762-2017

Operating Limits

  • Maximum allowable Pb content in condiments (GB 2762-2017): ≤1.0 mg/kg p.4 ↗
  • Maximum furnace program temperature: 2500 °C p.3 ↗