Induztr
Engineering Digest Document Viewer Related Models

Examining the Structural Influence of Site-Specific Phosphorylation by Ion Mobility Mass Spectrometry

Technical Reference Summary

Covered Models & Document Scope

Application note describing an IMS (ion mobility) LC/Q-TOF workflow for phosphopeptide analysis on the Agilent 6560 ion mobility LC/Q-TOF with Agilent Infinity UHPLC Nanodapter (G1988A), Infinity II 1290 binary pump, nanospray ion source (G1992A), and Agilent AssayMAP Bravo automated sample preparation; covers α- and β-casein digests and PhosphoMixes 1–3 Light Phosphopeptide Standards.

Technical Data

Parameter Value Condition
Mass range 100 to 1,700 m/z p.2 ↗ positive polarity, low (m/z 1700) mass range, SWARM autotune
Ion mobility mode mass range m/z 100 to 1,700 m/z p.3 ↗ IMS MS
High pressure funnel RF 100 V p.3 ↗
Trap funnel RF 100 V p.3 ↗
Drift tube exit voltage 250 V p.3 ↗
Column temperature 45 °C p.3 ↗
Autosampler temperature 4 °C p.3 ↗
Acquisition rate 10 spectra/s p.3 ↗ MS
Collision energy slope 2 p.3 ↗ MS, (Slope)*(m/z)/100+offset
Collision energy offset 3 p.3 ↗ MS
Isolation width Medium, m/z 4 m/z p.3 ↗
Max precursors per cycle 3 p.3 ↗
Threshold for MS/MS 1,000 counts p.3 ↗
Isotope model Peptides p.3 ↗
Drift tube length ~80 cm p.2 ↗
PhosphoMix dilution concentration 6.66 pmol/µL p.2 ↗ in 20 % acetonitrile, 0.1 % formic acid
Injection volume, digested α-casein 11 µL p.2 ↗ approximately 1 µg
Injection volume, digested β-casein 10 µL p.2 ↗ approximately 1 µg
Injection volume, phosphopeptide enrichment eluate 5 µL p.2 ↗ 1 µg of flowthrough
Injection volume, PhosphoMixes 2 µL p.2 ↗ 6.66 pmol

Operating Limits

  • Column temperature: 45 °C p.3 ↗
  • Autosampler temperature: 4 °C p.3 ↗
  • Mass range (low mass range tune): m/z 100 to 1,700 p.2 ↗