| Sample volume |
20 µL
p.3 ↗ |
glycoprotein or sialic acid standard |
| Sialic acid release |
2 hours at 80 °C
p.3 ↗ |
acid hydrolysis |
| DMB labeling |
3 hours at 50 °C
p.3 ↗ |
|
| LC system |
Agilent 1290 Infinity II LC System
p.3 ↗ |
|
| Autosampler |
Agilent Infinity multisampler (G7167B)
p.2 ↗ |
|
| Mass spectrometer |
Agilent 6545XT AdvanceBio LC/Q-TOF
p.3 ↗ |
|
| Ion source |
Dual AJS ESI
p.3 ↗ |
6545XT AdvanceBio LC/Q-TOF |
| Column |
Agilent InfinityLab PoroShell 120EC-C18
p.3 ↗ |
|
| Mobile phase A |
100 %
p.3 ↗ |
0.00 min, isocratic elution |
| Mobile phase B |
acetonitrile
p.3 ↗ |
|
| Flow rate |
0.4 mL/min
p.3 ↗ |
0.00–10.00 min |
| Run time |
10 min
p.2 ↗ |
rapid method |
| Gas temperature |
350 °C
p.3 ↗ |
6545XT AdvanceBio LC/Q-TOF |
| Sheath gas temperature |
400 °C
p.3 ↗ |
6545XT AdvanceBio LC/Q-TOF |
| Sheath gas flow |
12 L/min
p.3 ↗ |
6545XT AdvanceBio LC/Q-TOF |
| Nozzle voltage |
1,800 V
p.3 ↗ |
6545XT AdvanceBio LC/Q-TOF |
| Fragmentor voltage |
120 V
p.3 ↗ |
6545XT AdvanceBio LC/Q-TOF |
| Detection |
FLD or MS
p.3 ↗ |
|
| Software |
Agilent OpenLab CDS 2.3
p.2 ↗ |
|
| Kit order number |
GKK-407
p.6 ↗ |
AdvanceBio Sialic Acid profiling and quantitation kit |