Analyse von Restlösemitteln gemäß USP-Methode <467> auf dem
Technical Reference Summary
Covered Models & Document Scope
Application Note: analysis of residual solvents according to USP <467> using the Agilent 8890 GC system with 7697A headspace sampler, dual FID, and two columns: Agilent J&W DB-Select 624 UI for 467 (30 m × 0.32 mm, 1.8 µm, part no. 123-0334UI) and Agilent J&W HP-INNOWax.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Carrier gas | Helium, constant flow, 2 ml/min p.3 ↗ | |
| Inlet type | Split/Splitless p.3 ↗ | |
| Oven temperature | 85 °C p.3 ↗ | |
| Column 1 | J&W DB-Select 624 UI für 467, 30 m × 0.32 mm, 1.8 µm p.2 ↗ | part no. 123-0334UI |
| Column 2 | J&W HP-INNOWax p.2 ↗ | |
| Detector | dual flame ionization detectors (FID) p.2 ↗ | |
| Headspace sampler | Agilent 7697A p.2 ↗ | |
| Sample vials | clear 10-ml headspace vials with crimp cap p.2 ↗ | part no. 5190-2285 |
| Septa | Advanced Green, non-stick inlet septa p.2 ↗ | part no. 5183-4759-100 |
| Splitter | inert T-piece for Capillary Flow Technology p.2 ↗ | part no. G3184-60065 |
| Ferrules | short graphite ferrules for 0.1–0.32 mm columns, 10/pk p.2 ↗ | part no. 5080-8853 |
| Inlet liner | 2 mm, splitless, straight, deactivated p.2 ↗ | part no. 5181-8818 |
| Headspace transfer line / pre-CFT column | deactivated fused silica, 30 m × 0.25 mm ID × 0.35 mm OD p.2 ↗ | part no. 160-2255-30 |
Ordering Data
| Ordering Code | Part No. | Variant / Code | Applies To | Note |
|---|---|---|---|---|
| Sample vials p.2 ↗ | 5190-2285 | clear 10-ml headspace vials with crimp cap | ||
| Septa p.2 ↗ | 5183-4759-100 | Advanced Green, non-stick inlet septa | ||
| Splitter p.2 ↗ | G3184-60065 | inert T-piece for Capillary Flow Technology | ||
| Ferrules p.2 ↗ | 5080-8853 | short graphite ferrules for 0.1–0.32 mm columns, 10/pk | ||
| Ferrules p.2 ↗ | G3188-27502 | UltiMetal Plus, flexible metal, for 0.32 mm fused silica capillaries, 10/pk | ||
| Inlet liner p.2 ↗ | 5181-8818 | 2 mm, splitless, straight, deactivated | ||
| Headspace transfer line / pre-CFT column p.2 ↗ | 160-2255-30 | deactivated fused silica, 30 m × 0.25 mm ID × 0.35 mm OD | ||
| Column 1 p.2 ↗ | 123-0334UI | J&W DB-Select 624 UI für 467, 30 m × 0.32 mm, 1.8 µm |