| Temperature range |
+7 °C above ambient temperature to 100 °C °C
p.2 ↗ |
BF056-230V / BF056UL-120V |
| Temperature uniformity at 37°C |
0.2 ± K K
p.2 ↗ |
BF056-230V / BF056UL-120V |
| Temperature uniformity at 100°C |
1.0 ± K K
p.2 ↗ |
BF056-230V / BF056UL-120V |
| Temperature fluctuation at 37°C |
0.1 ± K K
p.2 ↗ |
BF056-230V / BF056UL-120V |
| Temperature fluctuation at 100°C |
0.2 ± K K
p.2 ↗ |
BF056-230V / BF056UL-120V |
| Heating up time to 100°C |
35 min
p.2 ↗ |
BF056-230V / BF056UL-120V |
| Heating up time to 37°C |
6 min
p.2 ↗ |
BF056-230V / BF056UL-120V |
| Recovery time after door was opened for 30 s at 37°C |
3 min
p.2 ↗ |
BF056-230V / BF056UL-120V |
| Rated voltage |
230 V
p.2 ↗ |
BF056-230V |
| Rated voltage |
120 V
p.2 ↗ |
BF056UL-120V |
| Power frequency |
50/60 Hz
p.2 ↗ |
BF056-230V |
| Power frequency |
60 Hz
p.2 ↗ |
BF056UL-120V |
| Nominal power |
0,4 kW
p.2 ↗ |
BF056-230V / BF056UL-120V |
| Unit fuse |
6,3 A
p.2 ↗ |
BF056-230V |
| Unit fuse |
12,5 A
p.2 ↗ |
BF056UL-120V |
| Phase (nominal voltage) |
1~
p.2 ↗ |
BF056-230V / BF056UL-120V |
| Interior volume |
59 L
p.2 ↗ |
BF056-230V / BF056UL-120V |
| Net weight of the unit (empty) |
39 kg
p.2 ↗ |
BF056-230V / BF056UL-120V |
| Load per rack |
30 kg
p.2 ↗ |
BF056-230V / BF056UL-120V |
| Permitted load |
70 kg
p.2 ↗ |
BF056-230V / BF056UL-120V |
| Wall clearance back |
160 mm
p.2 ↗ |
BF056-230V / BF056UL-120V |
| Wall clearance sidewise |
100 mm
p.2 ↗ |
BF056-230V / BF056UL-120V |
| Width net |
560 mm
p.2 ↗ |
BF056-230V / BF056UL-120V |
| Height net |
625 mm
p.2 ↗ |
BF056-230V / BF056UL-120V |
| Depth net |
565 mm
p.2 ↗ |
BF056-230V / BF056UL-120V |