Data Sheet Model CB 260
Technical Reference Summary
Covered Models & Document Scope
Model CB 260 CO2 incubator, interior volume 267 L. Four model versions: CB260-230V (standard, 200–230 V), CB260-230V-O (with O2 control 0.2–20%), CB260UL-120V (standard, 100–120 V, NEMA 5-20), CB260UL-120V-O (with O2 control 0.2–20%, 100–120 V).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Interior volume | 267 L p.3 ↗ | all models |
| Rated voltage | 200...230 V p.3 ↗ | CB260-230V / CB260-230V-O |
| Rated voltage | 100...120 V p.3 ↗ | CB260UL-120V / CB260UL-120V-O |
| Power frequency | 50/60 Hz p.3 ↗ | all models |
| Nominal power | 1,5 kW p.3 ↗ | all models |
| Unit fuse | 10,0 A p.3 ↗ | CB260-230V / CB260-230V-O |
| Unit fuse | 16,0 A p.3 ↗ | CB260UL-120V / CB260UL-120V-O |
| Phase (nominal voltage) | 1~ p.3 ↗ | all models |
| Temperature range | to 50 °C p.3 ↗ | all models |
| Temperature uniformity at 37°C | 0.4 ± K p.3 ↗ | all models |
| Temperature fluctuation at 37°C | 0.1 ± K p.3 ↗ | all models |
| Recovery time | 6 min p.3 ↗ | all models |
| Humidity range | 90 ...95 % RH p.3 ↗ | all models |
| CO2 range | 0 ...20 Vol.-% CO2 p.3 ↗ | all models |
| CO2 measuring technology | IR p.3 ↗ | all models |
| CO2 recovery time | 5 min p.3 ↗ | all models |
| Standard O2 control range | 0.2 ...20 Vol.-% O2 p.3 ↗ | CB260-230V-O / CB260UL-120V-O |
| Net weight of the unit (empty) | 125 kg p.3 ↗ | all models |
| Load per rack | 10 kg p.3 ↗ | all models |
| Permitted load | 40 kg p.4 ↗ | all models |
| Wall clearance back | 100 mm p.4 ↗ | all models |
| Wall clearance sidewise | 50 mm p.4 ↗ | all models |
| Width net | 740 mm p.4 ↗ | all models |
| Height net | 1,020 mm p.4 ↗ | all models |
| Depth net | 785 mm p.4 ↗ | all models |
Operating Limits
- Max temperature: 50 °C p.3 ↗
- Wall clearance back: 100 mm p.4 ↗
- Wall clearance sidewise: 50 mm p.4 ↗
- Permitted load: 40 kg p.4 ↗
- Load per rack: 10 kg p.3 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| 1-5 p.1 ↗ | CB260 | Model series CB, size 260 |
| 6-8 p.1 ↗ | UL | UL variant for 100–120 V / NEMA 5-20 (absent for 200–230 V CEE 7/7 variant) |
| 9-12 p.1 ↗ | 120V / 230V | Rated voltage variant: 120 V or 230 V |
| suffix p.1 ↗ | -O | With O2 control 0.2–20% (absent = standard without O2 control) |
Ordering Data
| Ordering Code | Part No. | Variant / Code | Applies To | Note |
|---|---|---|---|---|
| CB260-230V p.1 ↗ | 9040-0207 | 200...230 V 1~ 50/60 Hz -10,0 A, CEE 7/7, standard | ||
| CB260-230V-O p.1 ↗ | 9040-0208 | 200...230 V 1~ 50/60 Hz -10,0 A, CEE 7/7, with O2 control 0.2–20% | ||
| CB260UL-120V p.1 ↗ | 9040-0209 | 100...120 V 1~ 50/60 Hz -16,0 A, NEMA 5-20, standard | ||
| CB260UL-120V-O p.2 ↗ | 9040-0210 | 100...120 V 1~ 50/60 Hz -16,0 A, NEMA 5-20, with O2 control 0.2–20% |