CBF 260 CO2 incubators with hot air sterilization and humidity regulation, interior volume 267 L. Four model versions: CBF260-230V (Standard, 200...230 V, CEE 7/7 plug, Art.-No. 9040-0215), CBF260-230V-O (with O2 control 0.2–20%, Art.-No. 9040-0216), CBF260UL-120V (Standard, 100...120 V, NEMA 5-20 plug, Art.-No. 9040-0217), CBF260UL-120V-O (with O2 control 0.2–20%, Art.-No. 9040-0218).

Technical Data

Parameter Value Condition
Interior volume 267 L p.3 ↗ all models
Rated voltage 200...230 V p.3 ↗ CBF260-230V / CBF260-230V-O
Rated voltage 100...120 V p.3 ↗ CBF260UL-120V / CBF260UL-120V-O
Power frequency 50/60 Hz p.3 ↗ all models
Nominal power 1,5 kW p.3 ↗ all models
Unit fuse 10,0 A p.3 ↗ CBF260-230V / CBF260-230V-O
Unit fuse 16,0 A p.3 ↗ CBF260UL-120V / CBF260UL-120V-O
Phase (nominal voltage) 1~ p.3 ↗ all models
Temperature range to 50 °C p.3 ↗ all models
Temperature uniformity at 37°C 0.4 ± K p.3 ↗ all models
Temperature fluctuation at 37°C 0.1 ± K p.3 ↗ all models
Recovery time 6 min p.3 ↗ all models
Humidity range 50...95 % RH p.3 ↗ all models
CO2 range 0...20 Vol.-% CO2 p.3 ↗ all models
CO2 measuring technology IR p.3 ↗ all models
CO2 recovery time 5 min p.3 ↗ all models
Standard O2 control range 0.2...20 Vol.-% O2 p.3 ↗ CBF260-230V-O / CBF260UL-120V-O
Net weight of the unit (empty) 126 kg p.3 ↗ all models
Load per rack 10 kg p.3 ↗ all models
Permitted load 40 kg p.4 ↗ all models
Wall clearance back 100 mm p.4 ↗ all models
Wall clearance sidewise 50 mm p.4 ↗ all models
Width net 740 mm p.4 ↗ all models
Height net 1,020 mm p.4 ↗ all models
Depth net 785 mm p.4 ↗ all models

Operating Limits

  • max temperature: 50 °C p.3 ↗
  • wall clearance back: 100 mm p.4 ↗
  • wall clearance sidewise: 50 mm p.4 ↗
  • permitted load: 40 kg p.4 ↗
  • load per rack: 10 kg p.3 ↗

Type Code Decoder

Position Token Meaning
1-6 p.1 ↗ CBF260 Model series CBF, size 260
suffix p.1 ↗ UL UL version for 100...120 V supply with NEMA 5-20 plug
suffix p.1 ↗ -230V 200...230 V supply version with CEE 7/7 plug
suffix p.1 ↗ -120V 100...120 V supply version with NEMA 5-20 plug
suffix p.1 ↗ -O with O2 control 0.2–20%

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
CBF260-230V p.1 ↗ 9040-0215 200...230 V 1~ 50/60 Hz, 10,0 A, CEE 7/7, Standard
CBF260-230V-O p.1 ↗ 9040-0216 200...230 V 1~ 50/60 Hz, 10,0 A, CEE 7/7, with O2 control 0.2–20%
CBF260UL-120V p.1 ↗ 9040-0217 100...120 V 1~ 50/60 Hz, 16,0 A, NEMA 5-20, Standard
CBF260UL-120V-O p.2 ↗ 9040-0218 100...120 V 1~ 50/60 Hz, 16,0 A, NEMA 5-20, with O2 control 0.2–20%