| Temperature range |
+10 °C above ambient temperature to 300 °C
p.2 ↗ |
FP115-230V / FP115UL-120V |
| Temperature uniformity at 150°C |
1.4 ± K
p.2 ↗ |
FP115-230V / FP115UL-120V |
| Temperature fluctuation at 150°C |
0.3 ± K
p.2 ↗ |
FP115-230V / FP115UL-120V |
| Heating up time to 150°C |
19 min
p.2 ↗ |
FP115-230V / FP115UL-120V |
| Recovery time after door was opened for 30 s at 150°C |
5 min
p.2 ↗ |
FP115-230V / FP115UL-120V |
| Air change rate at 100°C |
32 x/h
p.2 ↗ |
FP115-230V / FP115UL-120V |
| Rated voltage |
230 V
p.2 ↗ |
FP115-230V |
| Rated voltage |
120 V
p.2 ↗ |
FP115UL-120V |
| Power frequency |
50/60 Hz
p.2 ↗ |
FP115-230V |
| Power frequency |
60 Hz
p.2 ↗ |
FP115UL-120V |
| Nominal power |
1,4 kW
p.2 ↗ |
FP115-230V / FP115UL-120V |
| Unit fuse |
10,0 A
p.2 ↗ |
FP115-230V |
| Unit fuse |
16,0 A
p.2 ↗ |
FP115UL-120V |
| Phase (nominal voltage) |
1~
p.2 ↗ |
FP115-230V / FP115UL-120V |
| Interior volume |
116 L
p.2 ↗ |
FP115-230V / FP115UL-120V |
| Net weight of the unit (empty) |
55 kg
p.2 ↗ |
FP115-230V / FP115UL-120V |
| Load per rack |
30 kg
p.2 ↗ |
FP115-230V / FP115UL-120V |
| Permitted load |
150 kg
p.2 ↗ |
FP115-230V / FP115UL-120V |
| Wall clearance back |
160 mm
p.2 ↗ |
FP115-230V / FP115UL-120V |
| Wall clearance sidewise |
100 mm
p.2 ↗ |
FP115-230V / FP115UL-120V |
| Width net |
714 mm
p.2 ↗ |
FP115-230V / FP115UL-120V |
| Height net |
732 mm
p.2 ↗ |
FP115-230V / FP115UL-120V |
| Depth net |
636 mm
p.2 ↗ |
FP115-230V / FP115UL-120V |
| Sound-pressure level |
43 dB(A)
p.3 ↗ |
FP115-230V / FP115UL-120V |
| Number of shelves (std./max.) |
2/5
p.3 ↗ |
FP115-230V / FP115UL-120V |