Induztr
Engineering Digest Document Viewer Related Models

Data Sheet Model KBF 470

Technical Reference Summary

Covered Models & Document Scope

KBF 470 constant climate chamber, two mains variants: KBF470-230V (Art. 9020-0480, 220...240 V 1~ 50/60 Hz, CEE 7/7 plug) and KBF470UL-120V (Art. 9020-0496, 120 V 1~ 50/60 Hz, NEMA 5-20 plug); both Standard option model.

Technical Data

Parameter Value Condition
Temperature range 0...70 °C p.2 ↗ KBF470-230V / KBF470UL-120V
Temperature range with humidity 10...70 °C p.2 ↗ KBF470-230V / KBF470UL-120V
Humidity range 10...80 % RH p.2 ↗ KBF470-230V / KBF470UL-120V
Max. heat compensation at 40°C 530 W p.2 ↗ KBF470-230V / KBF470UL-120V
Temperature uniformity at 25°C and 60% RH 0.2 ± K p.2 ↗ KBF470-230V / KBF470UL-120V
Temperature uniformity at 40°C and 75% RH 0.2 ± K p.2 ↗ KBF470-230V / KBF470UL-120V
Temperature fluctuation at 25°C and 60% RH 0.1 ± K p.2 ↗ KBF470-230V / KBF470UL-120V
Temperature fluctuation at 40°C and 75% RH 0.1 ± K p.2 ↗ KBF470-230V / KBF470UL-120V
Humidity fluctuation at 25°C and 60% RH 1.5 ± % RH p.2 ↗ KBF470-230V / KBF470UL-120V
Humidity fluctuation at 40°C and 75% RH 1.5 ± % RH p.2 ↗ KBF470-230V / KBF470UL-120V
Rated voltage 220...240 V p.2 ↗ KBF470-230V
Rated voltage 120 V p.2 ↗ KBF470UL-120V
Power frequency 50/60 Hz p.2 ↗ KBF470-230V / KBF470UL-120V
Nominal power 1,4 kW p.2 ↗ KBF470-230V / KBF470UL-120V
Unit fuse 16,0 A p.2 ↗ KBF470-230V / KBF470UL-120V
Phase (nominal voltage) 1~ p.2 ↗ KBF470-230V / KBF470UL-120V
Interior volume 472 L p.2 ↗ KBF470-230V / KBF470UL-120V
Net weight of the unit (empty) 225 kg p.2 ↗ KBF470-230V / KBF470UL-120V
Load per rack 40 kg p.2 ↗ KBF470-230V / KBF470UL-120V
Permitted load 200 kg p.2 ↗ KBF470-230V / KBF470UL-120V
Sound-pressure level at 25°C and 60% RH 51 dB(A) p.3 ↗ KBF470-230V / KBF470UL-120V
Energy consumption at 25°C and 60% RH 265 Wh/h p.3 ↗ KBF470-230V
Energy consumption at 25°C and 60% RH 290 Wh/h p.3 ↗ KBF470UL-120V
Width net 930 mm p.3 ↗ KBF470-230V / KBF470UL-120V
Height net 1,951 mm p.3 ↗ KBF470-230V / KBF470UL-120V

Operating Limits

  • Temperature range: 0...70 °C p.2 ↗
  • Temperature range with humidity: 10...70 °C p.2 ↗
  • Humidity range: 10...80 % RH p.2 ↗
  • Max. heat compensation at 40°C: 530 W p.2 ↗
  • Permitted load: 200 kg p.2 ↗
  • Load per rack: 40 kg p.2 ↗

Type Code Decoder

Position Token Meaning
1-3 p.1 ↗ KBF Model series: constant climate chamber with large temperature/humidity range
4-6 p.1 ↗ 470 Model size (nominal interior volume class, 472 L)
suffix p.1 ↗ UL UL variant for 120 V / NEMA 5-20 supply; absent = standard 220...240 V / CEE 7/7
suffix p.1 ↗ 230V Mains variant 220...240 V 1~ 50/60 Hz
suffix p.1 ↗ 120V Mains variant 120 V 1~ 50/60 Hz

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
KBF470-230V p.1 ↗ 9020-0480 220...240 V 1~ 50/60 Hz, CEE 7/7 plug, Standard Standard model version
KBF470UL-120V p.1 ↗ 9020-0496 120 V 1~ 50/60 Hz, NEMA 5-20 plug, Standard UL model version