Induztr
Engineering Digest Document Viewer Related Models

Data Sheet Model MKF CO₂ 1020

Technical Reference Summary

Covered Models & Document Scope

Model MKF 1020 CO₂ dynamic climate chamber for rapid temperature changes with humidity control. Two mains variants: MKF1020O2-400V (400 V 3~/N/PE 50 Hz, Art.-No. 9020-0528) and MKF1020-480V (480 V 3~/N/PE 60 Hz, Art.-No. 9020-0530).

Technical Data

Parameter Value Condition
Temperature range -48...180 °C p.3 ↗ MKF1020O2-400V, 50 Hz
Temperature range -48...180 °C p.3 ↗ MKF1020-480V, 60 Hz
Temperature uniformity dependent on set 0.1...1.0 ± K p.3 ↗ MKF1020O2-400V, 50 Hz
Temperature fluctuation dependent on set 0.1...0.2 ± K p.3 ↗ MKF1020O2-400V, 50 Hz
Cooling down time from 180°C to -40°C 70 min p.3 ↗ MKF1020O2-400V, 50 Hz
Temperature range with humidity 10...95 °C p.3 ↗ MKF1020O2-400V, 50 Hz
Humidity range 10...98 % RH p.3 ↗ MKF1020O2-400V, 50 Hz
Max. heat compensation at 25°C and 90% RH 1,400 W p.3 ↗ MKF1020O2-400V, 50 Hz
Rated voltage 400 V p.3 ↗ MKF1020O2-400V, 50 Hz
Rated voltage 480 V p.3 ↗ MKF1020-480V, 60 Hz
Power frequency 50 Hz p.3 ↗ MKF1020O2-400V
Power frequency 60 Hz p.3 ↗ MKF1020-480V
Nominal power 11,5 kW p.3 ↗ MKF1020O2-400V, 50 Hz
Unit fuse 32,0 A p.3 ↗ MKF1020O2-400V, 50 Hz
Phase (nominal voltage) 3~/N/PE p.3 ↗ MKF1020O2-400V, 50 Hz
Interior volume 991 L p.3 ↗ MKF1020O2-400V, 50 Hz
Net weight of the unit (empty) 570 kg p.3 ↗ MKF1020O2-400V, 50 Hz
Net weight of the unit (empty) 580 kg p.3 ↗ MKF1020-480V, 60 Hz
Load per rack 40 kg p.3 ↗ MKF1020O2-400V, 50 Hz
Permitted load 200 kg p.3 ↗ MKF1020O2-400V, 50 Hz
Width net 1,580 mm p.4 ↗ MKF1020O2-400V, 50 Hz
Height net 2,005 mm p.4 ↗ MKF1020O2-400V, 50 Hz
Depth net 1,480 mm p.4 ↗ MKF1020O2-400V, 50 Hz
Sound-pressure level 60 dB(A) p.4 ↗ MKF1020O2-400V, 50 Hz
Energy consumption at 25°C and 60% RH 3,000 Wh/h p.4 ↗ MKF1020O2-400V, 50 Hz

Operating Limits

  • Temperature range: -48...180 °C p.3 ↗
  • Temperature range with humidity: 10...95 °C p.3 ↗
  • Humidity range: 10...98 % RH p.3 ↗
  • Permitted load: 200 kg p.3 ↗
  • Load per rack: 40 kg p.3 ↗
  • Wall clearance back: 300 mm p.4 ↗

Type Code Decoder

Position Token Meaning
1-6 p.2 ↗ MKF1020 Model series MKF CO₂, size 1020
suffix p.2 ↗ O2 Model version suffix (only on 400 V 50 Hz variant)
voltage field p.2 ↗ 400V Rated voltage 400 V, 3~/N/PE, 50 Hz
voltage field p.2 ↗ 480V Rated voltage 480 V, 3~/N/PE, 60 Hz

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
MKF1020O2-400V p.2 ↗ 9020-0528 400 V 3~/N/PE 50 Hz, Standard version Standard plug
MKF1020-480V p.2 ↗ 9020-0530 480 V 3~/N/PE 60 Hz, Standard version NEMA L22-30P plug