Data Sheet Model MKF CO₂ 240

Technical Reference Summary

Covered Models & Document Scope

Model MKF 240 CO₂ dynamic climate chamber for rapid temperature changes with humidity control. Two variants: MKF240CO2-400V (Art.-No. 9020-0514, 400 V 3~/N/PE 50 Hz) and MKF240-480V (Art.-No. 9020-0516, 480 V 3~/N/PE 60 Hz).

Technical Data

Parameter Value Condition
Temperature range -50...180 °C p.3 ↗ MKF240CO2-400V, 50 Hz
Temperature range -50...180 °C p.3 ↗ MKF240-480V, 60 Hz
Temperature uniformity dependent on set 0.1...1.2 ± K p.3 ↗ MKF240CO2-400V, 50 Hz
Temperature fluctuation dependent on set 0.1...0.4 ± K p.3 ↗ MKF240CO2-400V, 50 Hz
Cooling down time from 180°C to -40°C 60 min p.3 ↗ MKF240CO2-400V, 50 Hz
Temperature range with humidity 10...95 °C p.3 ↗ MKF240CO2-400V, 50 Hz
Humidity range 10...98 % RH p.3 ↗ MKF240CO2-400V, 50 Hz
Max. heat compensation at 25°C and 90% RH 900 W p.3 ↗ MKF240CO2-400V, 50 Hz
Rated voltage 400 V p.3 ↗ MKF240CO2-400V, 50 Hz
Rated voltage 480 V p.3 ↗ MKF240-480V, 60 Hz
Power frequency 50 Hz p.3 ↗ MKF240CO2-400V
Power frequency 60 Hz p.3 ↗ MKF240-480V
Nominal power 6,8 kW p.3 ↗ MKF240CO2-400V, 50 Hz
Unit fuse 16,0 A p.3 ↗ MKF240CO2-400V, 50 Hz
Phase (nominal voltage) 3~/N/PE p.3 ↗ MKF240CO2-400V, 50 Hz
Interior volume 228 L p.3 ↗ MKF240CO2-400V, 50 Hz
Net weight of the unit (empty) 327 kg p.3 ↗ MKF240CO2-400V, 50 Hz
Net weight of the unit (empty) 335 kg p.3 ↗ MKF240-480V, 60 Hz
Load per rack 30 kg p.3 ↗ MKF240CO2-400V, 50 Hz
Permitted load 70 kg p.3 ↗ MKF240CO2-400V, 50 Hz
Width net 1,115 mm p.4 ↗ MKF240CO2-400V, 50 Hz
Height net 1,713 mm p.4 ↗ MKF240CO2-400V, 50 Hz
Depth net 925 mm p.4 ↗ MKF240CO2-400V, 50 Hz
Sound-pressure level 60 dB(A) p.4 ↗ MKF240CO2-400V, 50 Hz
Energy consumption at 25°C and 60% RH 1,400 Wh/h p.4 ↗ MKF240CO2-400V, 50 Hz

Operating Limits

  • Temperature range: -50...180 °C p.3 ↗
  • Temperature range with humidity: 10...95 °C p.3 ↗
  • Humidity range: 10...98 % RH p.3 ↗
  • Wall clearance back: 300 mm p.4 ↗
  • Wall clearance sidewise: 200 mm p.4 ↗
  • Permitted load: 70 kg p.3 ↗

Type Code Decoder

Position Token Meaning
1-3 p.1 ↗ MKF Series: dynamic climate chamber with humidity control
4-6 p.1 ↗ 240 Model size (interior volume class, 228 L)
suffix p.1 ↗ CO2 Refrigerant R744 (natural CO₂)
voltage/frequency field p.2 ↗ 400V 400 V 3~/N/PE, 50 Hz variant (Art.-No. 9020-0514)
voltage/frequency field p.2 ↗ 480V 480 V 3~/N/PE, 60 Hz variant (Art.-No. 9020-0516)

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
MKF240CO2-400V p.2 ↗ 9020-0514 400 V 3~/N/PE, 50 Hz, Standard version Main unit variant
MKF240-480V p.2 ↗ 9020-0516 480 V 3~/N/PE, 60 Hz, Standard version Main unit variant