Data Sheet Model MKFT 720

Technical Reference Summary

Covered Models & Document Scope

Model MKFT 720 dynamic climate chamber, 734 L interior volume. Two variants: MKFT720-400V (400 V 3~/N/PE 50 Hz, standard, Art.-No. 9020-0384) and MKFT720-480V-C (480 V 3~ 60 Hz, with voltage and frequency converter, Art.-No. 9020-0360).

Technical Data

Parameter Value Condition
Temperature range -70...180 °C p.3 ↗ MKFT720-400V
Temperature range -70...180 °C p.3 ↗ MKFT720-480V-C
Temperature uniformity dependent on set 0.3...2.0 ± K p.3 ↗ MKFT720-400V
Temperature fluctuation dependent on set 0.1...0.5 ± K p.3 ↗ MKFT720-400V
Cooling down time from 180°C to -70°C 120 min p.3 ↗ MKFT720-400V
Temperature range with humidity 10...95 °C p.3 ↗ MKFT720-400V
Humidity range 10...98 % RH p.3 ↗ MKFT720-400V
Max. heat compensation at 25°C and 90% RH 800 W p.3 ↗ MKFT720-400V
Rated voltage 400 V p.3 ↗ MKFT720-400V
Rated voltage 480 V p.3 ↗ MKFT720-480V-C
Power frequency 50 Hz p.3 ↗ MKFT720-400V
Power frequency 60 Hz p.3 ↗ MKFT720-480V-C
Nominal power 11 kW p.3 ↗ MKFT720-400V
Unit fuse 32,0 A p.3 ↗ MKFT720-400V
Phase (nominal voltage) 3~/N/PE p.3 ↗ MKFT720-400V
Phase (nominal voltage) 3~ p.3 ↗ MKFT720-480V-C
Interior volume 734 L p.3 ↗ MKFT720-400V
Net weight of the unit (empty) 635 kg p.3 ↗ MKFT720-400V
Net weight of the unit (empty) 887 kg p.3 ↗ MKFT720-480V-C
Load per rack 40 kg p.3 ↗ MKFT720-400V
Permitted load 160 kg p.3 ↗ MKFT720-400V
Sound-pressure level 69 dB(A) p.4 ↗ MKFT720-400V
Sound-pressure level 72 dB(A) p.4 ↗ MKFT720-480V-C
Energy consumption at 25°C and 60% RH 2,200 Wh/h p.4 ↗ MKFT720-400V
Number of shelves (std./max.) 1/11 p.4 ↗ MKFT720-400V

Operating Limits

  • Temperature range: -70...180 °C p.3 ↗
  • Temperature range with humidity: 10...95 °C p.3 ↗
  • Humidity range: 10...98 % RH p.3 ↗
  • Wall clearance back: 300 mm p.4 ↗
  • Wall clearance sidewise: 200 mm p.4 ↗
  • Permitted load: 160 kg p.3 ↗

Type Code Decoder

Position Token Meaning
1-5 p.1 ↗ MKFT7 Series MKFT, size 720
suffix p.1 ↗ 400V Standard variant, 400 V 3~/N/PE 50 Hz
suffix p.2 ↗ 480V-C Variant with voltage and frequency converter, 480 V 3~ 60 Hz

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
MKFT720-400V p.1 ↗ 9020-0384 400 V 3~/N/PE 50 Hz, standard Standard model
MKFT720-480V-C p.2 ↗ 9020-0360 480 V 3~ 60 Hz, with voltage and frequency converter Converter variant