Agilent CrossLab CS Electronic Leak Detector System

Technical Reference Summary

Covered Models & Document Scope

Agilent CrossLab CS Electronic Leak Detector cartridge system; probe filter replacement part G6694-60005; covers GC, GC/MS, ICP-OES and ICP-MS leak-checking applications

Technical Data

Parameter Value Condition
Gas rate for hydrogen 0.0025 mL/min p.3 ↗ filled bar indicator
Gas rate for helium 0.003 mL/min p.3 ↗ filled bar indicator
Gas rate for methane 0.014 mL/min p.3 ↗ filled bar indicator
Gas rate for nitrogen 1.0 mL/min p.3 ↗ empty bar indicator
Gas rate for carbon dioxide 0.03 mL/min p.3 ↗ empty bar indicator
Gas rate for helium (ICP application) 142.2 mL/min p.3 ↗ warning threshold
Gas rate for xenon 5.2 mL/min p.3 ↗
Probe filter replacement part number G6694-60005 p.10 ↗

Operating Limits

  • Probe filter backflush / replacement interval trigger: error "No Fltr Date" p.10 ↗
  • Power-off hold time for reboot: 3 seconds p.10 ↗
  • Wait time before power-on after reboot: 10 to 30 seconds p.10 ↗
  • Ambient-air baseline reset hold time: 2 seconds p.10 ↗